Performance of the TRUE dilatation balloon valvuloplasty catheter beyond rated burst pressure: A bench study

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Implications of hydrodynamic testing to guide sizing of self-expanding transcatheter heart valves for valve-in-valve procedures

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Intravascular iliac artery lithotripsy to enable transfemoral thoracic endovascular aortic repair

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Vascular function in adults with cyanotic congenital heart disease

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Implications of hydrodynamic testing to guide sizing of self-expanding transcatheter heart valves for valve-in-valve procedures

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Early and late risk of ischemic stroke after TAVR as compared to a nationwide background population

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. A novel supra-annular plane to predict TAVI prosthesis anchoring in raphe-type bicuspid aortic valve disease: the LIRA plane

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. Cardiovascular complications in patients with total cavopulmonary connection: A nationwide cohort study

    Research output: Contribution to journalJournal articleResearchpeer-review

  • Janarthanan Sathananthan
  • Mark Hensey
  • Stephanie Sellers
  • Aaron M Barlow
  • Adnan K Chhatriwalla
  • Keith B Allen
  • Anson Cheung
  • Lars Søndergaard
  • Philipp Blanke
  • Jian Ye
  • Jonathan Leipsic
  • David Wood
  • John Webb
View graph of relations

OBJECTIVES: We undertook an independent bench test assessing the performance of the TRUE dilatation (TD) balloon valvuloplasty catheter (Bard Vascular Inc., Tempe, AZ) beyond its rated burst pressure (RBP).

BACKGROUND: The TD balloon has a RBP of six atmospheres (atm), and its performance beyond this RBP is poorly understood. Techniques such as bioprosthetic valve fracture require inflation pressures beyond manufacturer recommendations.

METHODS: A 20, 22, 24, 26, and 28 mm TD balloon were inflated to increasing pressures in increments of 3 atm until balloon failure. Measurements were performed at the proximal, middle, and distal balloon segments with scientific digital calipers. Z-MED balloons (Braun Interventional Systems Inc., Bethlehem, PA) were used as a comparator.

RESULTS: The mean diameter at the middle of the 20, 22, 24, 26, and 28 mm TD balloon at nominal pressure (3 atm) was 20.02 ± 0.09, 21.77 ± 0.07, 23.9 ± 0.06, 25.82 ± 0.08, and 27.62 ± 0.08 mm, respectively. The maximal mean diameter at the middle of the 20, 22, 24, 26, and 28 mm TD balloon was 20.39 ± 0.03 mm (15 atm), 22.35 ± 0.03 mm (15 atm), 24.55 ± 0.02 mm (15 atm), 26.48 ± 0.02 mm (12 atm), and 28.39 ± 0.03 mm (12 atm), respectively. The 20/22/24 and 26/28 mm balloon failed when inflated beyond 15 atm and 12 atm, respectively. Failure was due to either leakage or longitudinal balloon rupture. TD balloons were more likely to maintain dimensions similar to their labeled size and less likely to fail at higher pressures as compared to Z-MED balloons.

CONCLUSION: The TD balloon catheter maintains a similar diameter to its labeled size, when inflated beyond its RBP. When inflated beyond 12 atm, the TD balloon can fail due to either leakage or rupture. This has implications for percutaneous structural heart interventions.

Original languageEnglish
JournalCatheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
Volume96
Issue number2
Pages (from-to)E187-E195
ISSN1522-1946
DOIs
Publication statusPublished - Aug 2020

    Research areas

  • aortic stenosis, balloon valvuloplasty, degenerative valve, other, valve restenosis

ID: 59143434