Implications of hydrodynamic testing to guide sizing of self-expanding transcatheter heart valves for valve-in-valve procedures

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Performance of the TRUE dilatation balloon valvuloplasty catheter beyond rated burst pressure: A bench study

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Intravascular iliac artery lithotripsy to enable transfemoral thoracic endovascular aortic repair

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Vascular function in adults with cyanotic congenital heart disease

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Performance of the TRUE dilatation balloon valvuloplasty catheter beyond rated burst pressure: A bench study

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Early and late risk of ischemic stroke after TAVR as compared to a nationwide background population

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. A novel supra-annular plane to predict TAVI prosthesis anchoring in raphe-type bicuspid aortic valve disease: the LIRA plane

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. Cardiovascular complications in patients with total cavopulmonary connection: A nationwide cohort study

    Research output: Contribution to journalJournal articleResearchpeer-review

  • Janarthanan Sathananthan
  • Mark Hensey
  • Rob Fraser
  • Uri Landes
  • Philipp Blanke
  • Hoda Hatoum
  • Lakshmi Prasad Dasi
  • Alexander Sedaghat
  • Vinayak N Bapat
  • Jonathon Leipsic
  • Lars Søndergaard
  • David A Wood
  • John G Webb
View graph of relations

AIMS: The commonly used valve-in-valve (VIV) app recommends sizing based on dimensions of both the transcatheter heart valve (THV) and bioprosthetic surgical valve. The implications of hydrodynamic testing to guide VIV sizing are poorly understood. This bench study assessed the hydrodynamic performance of different sizes of self-expanding supra-annular THVs in three different surgical aortic bioprostheses at different implantation depths.

METHODS: A small versus medium ACURATE neo (ACn), and a 26 mm versus 29 mm Evolut R were assessed after VIV implantation in 25 mm Mitroflow, Mosaic, and Magna Ease aortic surgical bioprostheses, at three implantation depths (+2 mm, -2 mm, and -6 mm).

RESULTS: The medium-sized ACn had lower gradients compared to the small ACn when the THV was implanted high (+2 mm, or -2 mm). The 29 mm Evolut R had lower gradients compared to a 26 mm Evolut R for all implantation depths, except for a depth of -2 mm in the 25 mm Mitroflow. The medium ACn and 29 mm Evolut R had larger effective orifice areas compared to the small ACn and 26 mm Evolut R, respectively. Both Evolut R sizes had acceptable regurgitant fractions (<15%), while both ACn sizes were above the acceptable performance criteria (>15%), at all implantation depths.

CONCLUSIONS: Use of a larger self-expanding THV was associated with superior hydrodynamic performance if the THV was implanted high. Hydrodynamic testing can provide additional information to the VIV app to help guide VIV sizing.

Original languageEnglish
JournalCatheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
Volume96
Issue number3
Pages (from-to)E332-E340
ISSN1522-1946
DOIs
Publication statusPublished - 1 Sep 2020

    Research areas

  • ACURATE neo, hydrodynamic performance, self-expanding, transcatheter heart valve, valve-in-valve