Implications of hydrodynamic testing to guide sizing of self-expanding transcatheter heart valves for valve-in-valve procedures

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Impact of early hospital discharge on clinical outcomes after transcatheter aortic valve implantation

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. State-of-the-art preclinical testing of the OMEGATM left atrial appendage occluder

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Spontaneous thrombosis of a transcatheter aortic valve replacement-induced aortic root pseudoaneurysm

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Patient-Specific Implantation Technique to Obtain Neo-Commissural Alignment With Self-Expanding Transcatheter Aortic Valves

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Technical Considerations for Transcatheter Aortic Valve Replacement With the Navitor Transcatheter Heart Valve

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. First-in-human results of the OMEGA™ Left Atrial Appendage Occluder for Patients with Non-Valvular Atrial Fibrillation

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Impact of early hospital discharge on clinical outcomes after transcatheter aortic valve implantation

    Research output: Contribution to journalJournal articleResearchpeer-review

  • Janarthanan Sathananthan
  • Mark Hensey
  • Rob Fraser
  • Uri Landes
  • Philipp Blanke
  • Hoda Hatoum
  • Lakshmi Prasad Dasi
  • Alexander Sedaghat
  • Vinayak N Bapat
  • Jonathon Leipsic
  • Lars Søndergaard
  • David A Wood
  • John G Webb
View graph of relations

AIMS: The commonly used valve-in-valve (VIV) app recommends sizing based on dimensions of both the transcatheter heart valve (THV) and bioprosthetic surgical valve. The implications of hydrodynamic testing to guide VIV sizing are poorly understood. This bench study assessed the hydrodynamic performance of different sizes of self-expanding supra-annular THVs in three different surgical aortic bioprostheses at different implantation depths.

METHODS: A small versus medium ACURATE neo (ACn), and a 26 mm versus 29 mm Evolut R were assessed after VIV implantation in 25 mm Mitroflow, Mosaic, and Magna Ease aortic surgical bioprostheses, at three implantation depths (+2 mm, -2 mm, and -6 mm).

RESULTS: The medium-sized ACn had lower gradients compared to the small ACn when the THV was implanted high (+2 mm, or -2 mm). The 29 mm Evolut R had lower gradients compared to a 26 mm Evolut R for all implantation depths, except for a depth of -2 mm in the 25 mm Mitroflow. The medium ACn and 29 mm Evolut R had larger effective orifice areas compared to the small ACn and 26 mm Evolut R, respectively. Both Evolut R sizes had acceptable regurgitant fractions (<15%), while both ACn sizes were above the acceptable performance criteria (>15%), at all implantation depths.

CONCLUSIONS: Use of a larger self-expanding THV was associated with superior hydrodynamic performance if the THV was implanted high. Hydrodynamic testing can provide additional information to the VIV app to help guide VIV sizing.

Original languageEnglish
JournalCatheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
Volume96
Issue number3
Pages (from-to)E332-E340
ISSN1522-1946
DOIs
Publication statusPublished - 1 Sep 2020

    Research areas

  • ACURATE neo, hydrodynamic performance, self-expanding, transcatheter heart valve, valve-in-valve

ID: 59143769