Research
Print page Print page
Switch language
Rigshospitalet - a part of Copenhagen University Hospital
Published

Few adverse reactions to metal on metal articulation in total hip arthroplasty in a review study on 358 consecutive cases with 1 to 5 years follow-up

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Treatment Algorithm for Patients with Non-arthritic Hip Pain, Suspect for an Intraarticular Pathology

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Persistent wound drainage after tumor resection and endoprosthetic reconstruction of the proximal femur

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Yngre patienter med hofteledssmerter

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. The interface between periacetabular osteotomy, hip arthroscopy and total hip arthroplasty in the young adult hip

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Osseous osteoarthritic-like changes and joint mobility of the temporomandibular joints and upper cervical spine: is there a relation?

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations
The aim of the present study was to determine the frequency of adverse reaction to metal on metal total hip arthroplasty using a M2a-38 articulation and a Magnum articulation, (Biomet Warsaw, Indiana).All patients who had received a Metal on Metal bearing prosthesis, in two centres in Copenhagen, were asked to complete a questionnaire about groin pain. Patients with self-reported groin pain, 68/314, underwent a physical examination and had Co and Cr ion levels measured in full blood samples. Fifty patients also had a CT scan performed.The present study found one patient with bilateral arthroplasty, who had an adverse reaction on one side. In addition the study showed that females had higher values of Co and Cr, and that younger patient reported groin pain more often.The conclusion of this study is that the number of adverse reactions is low, despite the time of observation being relatively short, no high frequency of adverse reactions to this prosthesis is expected.
Original languageEnglish
JournalThe Open Orthopaedics Journal
Volume6
Pages (from-to)366-70
Number of pages5
ISSN1874-3250
DOIs
Publication statusPublished - 2012