Forskning
Udskriv Udskriv
Switch language
Rigshospitalet - en del af Københavns Universitetshospital
Udgivet

Visualizing spikes in source-space: Rapid and efficient evaluation of magnetoencephalography

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Steady-state visual evoked potential temporal dynamics reveal correlates of cognitive decline

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Diagnostic added value of electrical source imaging in presurgical evaluation of patients with epilepsy: A prospective study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Diagnostic yield of high-density versus low-density EEG: The effect of spatial sampling, timing and duration of recording

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Oscillatory connectivity as a diagnostic marker of dementia due to Alzheimer's disease

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  5. Automated EEG source imaging: A retrospective, blinded clinical validation study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Extensive astrocyte metabolism of γ-aminobutyric acid (GABA) sustains glutamine synthesis in the mammalian cerebral cortex

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Author Correction: Identification of epilepsy-associated neuronal subtypes and gene expression underlying epileptogenesis

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Identification of epilepsy-associated neuronal subtypes and gene expression underlying epileptogenesis

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Transcutaneous Vagus Nerve Stimulation in Patients With Severe Traumatic Brain Injury: A Feasibility Trial

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

OBJECTIVE: Reviewing magnetoencephalography (MEG) recordings is time-consuming: signals from the 306 MEG-sensors are typically reviewed divided into six arrays of 51 sensors each, thus browsing each recording six times in order to evaluate all signals. A novel method of reconstructing the MEG signals in source-space was developed using a source-montage of 29 brain-regions and two spatial components to remove magnetocardiographic (MKG) artefacts. Our objective was to evaluate the accuracy of reviewing MEG in source-space.

METHODS: In 60 consecutive patients with epilepsy, we prospectively evaluated the accuracy of reviewing the MEG signals in source-space as compared to the classical method of reviewing them in sensor-space.

RESULTS: All 46 spike-clusters identified in sensor-space were also identified in source-space. Two additional spike-clusters were identified in source-space. As 29 source-channels can be easily displayed simultaneously, MEG recordings had to be browsed only once. Yet, this yielded a global coverage of the recorded signals and enhanced detectability of epileptiform discharges because MKG-artefacts were suppressed and did not impede evaluation in source-space.

CONCLUSIONS: Our results show that reviewing MEG recordings in source-space is accurate and much more rapid than the classical method of reviewing in sensor-space.

SIGNIFICANCE: This novel method facilitates the clinical use of MEG.