Forskning
Udskriv Udskriv
Switch language
Rigshospitalet - en del af Københavns Universitetshospital
Udgivet

Transcatheter aortic valve implantation with the self-expanding Portico valve system in an all-comers population: procedural and clinical outcomes

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Mechanical circulatory support for decompensated heart failure: the last remaining indication for intra-aortic balloon pump?

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Transcatheter mitral valve repair: a step back from the edge

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Percutaneous coronary and structural interventions in women: a position statement from the EAPCI Women Committee

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. The invisible army of women in interventional cardiology: EAPCI Women mission to make them visible

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Mortality and morbidity of major congenital heart disease related to general prenatal screening for malformations

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Long-Term Risk of Infective Endocarditis After Transcatheter Aortic Valve Replacement

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Long-Term Outcomes of Partial Oral Treatment of Endocarditis

    Publikation: Bidrag til tidsskriftLetterForskning

Vis graf over relationer

AIMS: Transcatheter aortic valve implantation (TAVI) is an established therapy for patients with severe aortic stenosis (AS). The aim of this study was to evaluate the newer-generation Portico TAVI system in an all-comers population.

METHODS AND RESULTS: This single-centre study included 216 patients with severe AS (Society of Thoracic Surgeons [STS] score 4.3±3.0%). The Portico valve was implanted using the transfemoral (91.2%), transsubclavian (5.6%) and transcaval (3.2%) access. Device success was achieved in 94.4% of cases. At 30 days, mortality and stroke rates were 2.3% and 0.5%, respectively. Early safety was achieved in 91.7% of cases. More-than-mild paravalvular leak (PVL), as assessed by echocardiogram, was observed in 3.4% of the patients, with rates of 4.9% and 1.9% in the first and second half of the cohort, respectively. A permanent pacemaker was implanted in 15.8% of those without prior pacemaker, with a rate of 11.1% in the second half of the cohort. At one year, incidence rates for all-cause mortality and stroke were 12.3% and 2.3%, respectively. In the low-risk group (STS <4%; n=128), Kaplan-Meier estimates at 30 days and one year were 0% and 7.5% for all-cause mortality and 0.8% and 2.2% for stroke, respectively. Haemodynamic improvements persisted over time with a mean transvalvular gradient of 7.0±3.0 mmHg at one-year follow-up.

CONCLUSIONS: The Portico TAVI system was safe to implant and achieved a high device success rate. With learning curve effects, the device achieves lower rates of PVL and pacemaker implantation and provides adequate clinical and haemodynamic outcomes up to one year.