Forskning
Udskriv Udskriv
Switch language
Rigshospitalet - en del af Københavns Universitetshospital
Udgivet

Imaging risk features for device related pulmonary artery injury after left atrial appendage closure with Amplatzer™ Amulet™ device

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Impact of early hospital discharge on clinical outcomes after transcatheter aortic valve implantation

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. State-of-the-art preclinical testing of the OMEGATM left atrial appendage occluder

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Spontaneous thrombosis of a transcatheter aortic valve replacement-induced aortic root pseudoaneurysm

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Feasibility and safety of a fully percutaneous transcatheter aortic valve replacement program

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Patient-Specific Implantation Technique to Obtain Neo-Commissural Alignment With Self-Expanding Transcatheter Aortic Valves

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. First-in-human results of the OMEGA™ Left Atrial Appendage Occluder for Patients with Non-Valvular Atrial Fibrillation

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Impact of early hospital discharge on clinical outcomes after transcatheter aortic valve implantation

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Contemporary management of severe symptomatic bicuspid aortic valve stenosis: the BiTri Registry

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

OBJECTIVES: This study aimed to find imaging risk features for device related-pulmonary artery (PA) injury (DR-PAI) in patients after left atrial appendage closure (LAAC).

BACKGROUND: Cardiac tamponade resulting from DR-PAI is a rare but life-threatening complication of LAAC.

METHODS: In vitro analysis of Amplatzer™ Amulet™ (Abbott, MN) device was done. Measurements of the distance between PA and Amplatzer lobe at its middle part, distal part, and along the stabilizing wires' trajectory (wires-to-PA distance) were taken in 100 consecutive patients on post-LAAC computed tomography (CT) studies. Clinical outcomes were collected.

RESULTS: In vitro analysis indicated that both middle and distal lobe had to be in close PA proximity (<1.5 mm) for the stabilizing wires to cause risk for DR-PAI, configuration called ''cuddling lobe orientation''. On CT measurements middle lobe-PA distance was 4.7 mm (IQR = 2.7-9.5), with close proximity in 4 (4%) patients, and distal lobe-PA distance was 3.2 mm (IQR = 1.8-7.2 mm), with close proximity in 17 (17%) patients. Stabilizing wires were pointing toward PA in 47 patients (47%) with median wire-to-PA distance 5.7 mm (IQR = 3.6-8.5 mm). ''Cuddling'' was found in 2 (2%) patients and resulted in shorter wires-to-PA distance vs no ''cuddling'' group (2.3 vs 5.8 mm p <.01). At 2.9 ± 1.0 yrs of follow-up, the two patients with ''cuddling lobe orientation'' on post-LAAC CT scan developed late cardiac tamponades (p <.001). One of those required surgery, which confirmed DR-PAI.

CONCLUSIONS: ''Cuddling lobe orientation'' of Amulet device with the PA was associated with short wires-to-PA distance and late pericardial effusions, including DR-PAI. Hence, such device-to-PA configuration should be avoided.

OriginalsprogEngelsk
TidsskriftCatheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions
Vol/bind98
Udgave nummer3
Sider (fra-til)E420-E426
ISSN1522-1946
DOI
StatusUdgivet - sep. 2021