Evaluation of healthy muscle tissue by strain and shear wave elastography - Dependency on depth and ROI position in relation to underlying bone

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Validation of an assessment tool for estimation of abdominal aortic aneurysm compression in diagnostic ultrasound

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Vector velocity volume flow estimation: Sources of error and corrections applied for arteriovenous fistulas

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. First report on intraoperative vector flow imaging of the heart among patients with healthy and diseased aortic valves

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. In-vivo validation of fast spectral velocity estimation techniques

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Pressure Difference Estimation in Non-stenotic Carotid Bifurcation Phantoms Using Vector Flow Imaging

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Evaluation of 2D super-resolution ultrasound imaging of the rat renal vasculature using ex vivo micro-computed tomography

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Highlights of the development in ultrasound during the last 70 years: A historical review

    Publikation: Bidrag til tidsskriftReviewForskningpeer review

Vis graf over relationer

PURPOSE: The aim of this study was to evaluate the influence of depth and underlying bone on strain ratios and shear wave speeds for three different muscles in healthy volunteers. For strain ratios the influence from different reference region-of-interest positions was also evaluated.

MATERIAL AND METHODS: Ten healthy volunteers (five males and five females) had their biceps brachii, gastrocnemius, and quadriceps muscle examined with strain- and shear wave elastography at three different depths and in regions located above bone and beside bone. Strain ratios were averaged from cine-loops of 10s length, and shear wave speeds were measured 10 times at each target point. The distance from the skin surface to the centre of each region-of-interest was measured. Measurements were evaluated with descriptive statistics and linear regression.

RESULTS: Linear regression showed a significant influence on strain ratio measurements from the reference region-of-interest position, i.e. being above the same structures as the target region-of-interest or not (means: 1.65 and 0.78; (P<0.001)). For shear wave speeds, there was a significant influence from depth and location above or beside bone (P=0.011 and P=0.031).

CONCLUSION: Strain ratio values depend significantly on reference and target region-of-interest being above the same tissue, for instance bone. Strain ratios were not influenced by depth in this study. Shear wave speeds decreased with increasing scanning depth and if there was bone below the region-of-interest.

OriginalsprogEngelsk
TidsskriftUltrasonics
Vol/bind71
Sider (fra-til)127-33
Antal sider7
ISSN0041-624X
DOI
StatusUdgivet - sep. 2016