Abstract
OBJECTIVE: To investigate the feasibility of early sound processor loading with two separate bone-anchored hearing aid implants 1 to 2 weeks after implantation. Secondary endpoints included registration of any postoperative complications, skin reactions, and any adverse events.
STUDY DESIGN: Prospective, nonrandomized, noncontrolled, multicenter clinical trial.
SETTING: Tertiary referral center.
PATIENTS: Sixty adult patients, eligible for bone-anchored hearing aid surgery, were enrolled.
INTERVENTION: Implantation of the bone-anchored hearing aid implants were done using minimally invasive surgery with tissue preservation.
MAIN OUTCOME MEASURE: Implant stability quotient (ISQ) values were recorded using resonance frequency analysis. Follow-ups were scheduled 5 to 12 days, 4 to 8 weeks, 6 months, and 12 months postoperatively. Differences between the lowest ISQ values (ISQ low) over time were analyzed by paired-samples t test with a significance level of 0.05. Skin and soft tissue reactions were assessed according to Holger's classification.
RESULTS: Loading time of the Ponto Wide implant was 11.6 ± 1.4 days (SD) in average (median, 12 d; range, 7-19 d). Loading of the Ponto BHX implant was done 9.7 ± 3.5 days (SD) postoperatively (median, 9 d; range, 5-19 d). The implant stability increased significantly during the 12-month study period with both implants. One Ponto BHX implant was lost. Skin and soft tissue reactions were few in both groups. Across all planned follow-up visits, no skin and soft tissue reactions were observed in 86% (Ponto BHX implant) and 85% (Ponto Wide implant).
CONCLUSION: Very early loading of sound processors after percutaneous bone-anchored hearing system surgery is safe already 1 to 2 weeks postoperatively.
| Original language | English |
|---|---|
| Journal | Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology |
| Volume | 45 |
| Issue number | 5 |
| Pages (from-to) | 556-563 |
| Number of pages | 8 |
| ISSN | 1531-7129 |
| DOIs | |
| Publication status | Published - 1 Jun 2024 |
Keywords
- Adult
- Aged
- Aged, 80 and over
- Bone-Anchored Prosthesis
- Female
- Hearing Aids
- Humans
- Male
- Middle Aged
- Prospective Studies
- Prosthesis Implantation/methods
- Suture Anchors
- Treatment Outcome
- Young Adult
- Osseointegration
- Skin reaction
- Implant stability quotient
- Bone-anchored hearing system
- Holgers grading system