Validity and reliability of 3D US for the detection of erosions in patients with rheumatoid arthritis using MRI as the gold standard

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Lung Ultrasound in the Assessment of Pulmonary Complications After Lung Transplantation

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Elastografie in der Brustbildgebung

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Simulation-Based Training of Ultrasound-Guided Procedures in Radiology - A Systematic Review

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. The EFSUMB Guidelines and Recommendations for the Clinical Practice of Elastography in Non-Hepatic Applications: Update 2018

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. The Use of Handheld Ultrasound Devices - An EFSUMB Position Paper

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Perfusion in bone marrow lesions assessed on DCE-MRI and its association with pain in knee osteoarthritis: a cross-sectional study

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Associations between shoulder symptoms and concomitant pathology in patients with traumatic supraspinatus tears

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Early development of tendinopathy in humans: Sequence of pathological changes in structure and tissue turnover signaling

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

PURPOSE: To test the reliability and validity of a 3D US erosion score in RA using MRI as the gold standard.

MATERIALS AND METHODS: RA patients were examined with 3D US and 3 T MRI over the 2nd and 3rd metacarpophalangeal joints. 3D blocks were evaluated by two investigators. The erosions were estimated according to a semi-quantitative score (SQS) (0 - 3) and a quantitative score (QS) (mm²). MRI was evaluated according to the RAMRIS score. For the estimation of reliability, intra-class correlation coefficients (ICC) were used. Validity was tested using Spearman's rho (rs). The sensitivity and specificity were also calculated.

RESULTS: 28 patients with RA were included. The ICC for the inter-observer reliability in the QS was 0.41 and 0.13 for the metacarpal bone and phalangeal bone, respectively, and 0.86 and 0.16, respectively, in the SQS.  The ICC for the intra-observer reliability in the QS was 0.75 and 0.48 for the metacarpal bone and phalangeal bone, respectively, and 0.83 and 0.60, respectively, in the SQS.  The correlation with MRI for the metacarpal bone was significant, with values of 0.73 (p = 0.0001) (SQ) and 0.74 (p = 0.0001) (SQS). For the phalangeal bone, bad correlation was found: 0.28 (p = 0.145) (SQ) and 0.26 (p = 0.57) (SQS). The sensitivity and specificity for the metacarpal bone were 86 % and 85 % respectively. For the phalangeal bone they were 60 % and 97 %, respectively.

CONCLUSION: Good inter- and intra-observer reliability and correlation with MRI were seen in the assessment of erosions with 3D US in the metacarpal bone, while the results were low and insignificant for the phalangeal bone, indicating that 3D US still has room for improvement.

Original languageEnglish
JournalUltraschall in der Medizin
Volume35
Issue number2
Pages (from-to)137-41
Number of pages5
ISSN0172-4614
DOIs
Publication statusPublished - Apr 2014

ID: 44650193