Validation of an assessment tool for estimation of abdominal aortic aneurysm compression in diagnostic ultrasound

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Vector velocity volume flow estimation: Sources of error and corrections applied for arteriovenous fistulas

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. First report on intraoperative vector flow imaging of the heart among patients with healthy and diseased aortic valves

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Internal strain estimation for quantification of human heel pad elastic modulus: A phantom study

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. In-vivo validation of fast spectral velocity estimation techniques

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Automatic and Objective Assessment of Motor Skills Performance in Flexible Bronchoscopy

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Endovascular Treatment of Chronic and Acute on Chronic Mesenteric Ischaemia: Results From a National Cohort of 245 Cases

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Learning Curves and Competences of Vascular Trainees Performing Open Aortic Repair in a Simulation-Based Environment

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Three and two dimensional ultrasound is as accurate as computed tomography in aortic sac assessment after endovascular aortic repair

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. Simulation Based Training and Assessment in Open Vascular Surgery: A Systematic Review

    Research output: Contribution to journalReviewpeer-review

View graph of relations

The study investigated ultrasound (US) transducer push, tantamount to applied transducer pressure, during abdominal aortic aneurysm (AAA) US scanning in a simulated non-clinical setup. During an assessment of maximal AAA diameter on a three-dimensional print-based AAA phantom, US transducer push varied as much as 2000% (range: 0.52-12.45 kPa) amongst 16 experienced sonographers. The mean transducer push was 5.54 ± 3.91 kPa (CV = 0.71). Deformation of a standardized gel-pad allowed for transducer push calculation based on US images; Young's modulus of the gel-pad was estimated to 44,26 N/m2. The method is theoretically validated in a safe and non-clinical environment. Future investigations with the aim of clinical validation of the gel-pad principle on AAA patients are suggested, including the objectification of the magnitude of an eventual transducer push-related error during US AAA diameter measurement.

Original languageEnglish
JournalUltrasonics
Volume116
Pages (from-to)106484
ISSN0041-624X
DOIs
Publication statusE-pub ahead of print - 29 May 2021