Three and two dimensional ultrasound is as accurate as computed tomography in aortic sac assessment after endovascular aortic repair

3 Citations (Scopus)

Abstract

BACKGROUND: To compare aortic sac changes after endovascular aneurysm repair (EVAR) assessed by three-dimensional ultrasound (3D-US), two-dimensional ultrasound (2D-US), and traditional computed tomographic angiography (CTA).

METHODS: Using volume assessment with three-dimensional CTA (3D-CTA-volume) as the gold standard, this study investigated aortic sac changes at three and 12 months after EVAR with three different ultrasound methods (2D-US anterior-posterior (AP) diameter, 3D-US AP centerline diameter, and 3D-US partial volume), and traditional CT multiplanar outer-to-outer diameter (CT-MPR OTO diameter). From august 1st, 2011 to January 2014, consecutive EVAR patients (n = 113) were available for analysis in two time intervals; 1) between preoperative and three-month follow-up and 2) between three and 12 month follow-up.

RESULTS: The risk of missing true aortic sac growth (false negative finding) at three-month postoperative visit using 3D-US partial volume, 3D-US AP centerline diameter, 2D-US AP diameter, and CT-MPR OTO diameter was 19%, 21%, 22%, and 18%, respectively. Corresponding low sensitivities (0% to 21%) and kappa-values (<0.50) in detecting aortic sac changes were found. The risk of missing true growth between three and 12 months were lower (6%, 5%, 6%, and 6%, respectively), and matching sensitivities 33%, 33%, 17%, and 17%, respectively.

CONCLUSIONS: All tested methods for aortic sac changes were as good as traditional CT-MPR OTO diameter and corresponded poorly with 3D-CTA-volume at three months postoperative visit but substantially better after 12 months where the residual sac change was more profound.

Original languageEnglish
JournalAnnals of Vascular Surgery
Volume72
Pages (from-to)321-329
Number of pages9
ISSN0890-5096
DOIs
Publication statusPublished - Apr 2021

Fingerprint

Dive into the research topics of 'Three and two dimensional ultrasound is as accurate as computed tomography in aortic sac assessment after endovascular aortic repair'. Together they form a unique fingerprint.

Cite this