Simultaneous quantification of high-dose naloxone and naloxone-3-β-d-glucuronide in human plasma by UHPLC-MS/MS

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Fluispotter, a novel automated and wearable device for accurate volume serial dried blood spot sampling

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. The influence of sample collection methodology and sample preprocessing on the blood metabolic profile

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Development of a fast method for direct analysis of intact synthetic insulins in human plasma: the large peptide challenge

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Current strategies in the discovery of small-molecule biomarkers for Alzheimer's disease

    Research output: Contribution to journalReviewResearchpeer-review

  5. First example of hepatocyte transplantation to alleviate ornithine transcarbamylase deficiency, monitored by NMR-based metabonomics

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

AIM: High-dose administration of the μ-opioid receptor inverse agonist naloxone (NX), has previously been demonstrated to reinstate nocifensive behavior in the late phase of inflammatory injuries. However, no current analytical methods can provide pharmacokinetic insight into the pharmacodynamic response of high-dose administration of NX.

MATERIALS & METHODS: Based on protein precipitation using 50 μl human plasma, NX and naloxone-β-d-glucuronide (NXG) was analysed by UHPLC-MS/MS with 6 min cycle time. Results: A method for quantification of high-dose administered NX and NXG was developed and validated with intra- and interday precision and accuracy within ≤8.5% relative standard deviation (RSD) and -1.2-5.5% relative error (RE) for NX and ≤9.6% RSD and 0.6-6.5% RE for NXG. The method show excellent internal standard corrected matrix effects.

CONCLUSION: A rapid UHPLC-MS/MS method was developed for quantification of NX and NXG in human plasma within 10-4000 ng/ml.

Original languageEnglish
JournalBioanalysis
Volume11
Issue number3
Pages (from-to)165-173
Number of pages9
ISSN1757-6180
DOIs
Publication statusPublished - 2020

    Research areas

  • high-dose, naloxone, naloxone-3-β-D-glucuronide, quantification, UHPLC-MS/MS