Skip to main navigation Skip to search Skip to main content

Short-term repeatability of myocardial blood flow using 82Rb PET/CT: The effect of arterial input function position and motion correction

Yuka Otaki, Martin Lyngby Lassen, Osamu Manabe, Evann Eisenberg, Heidi Gransar, Frances Wang, Yoon Jae Lee, Evangelos Tzolos, Daniel S Berman, Piotr J Slomka

15 Citations (Scopus)

Abstract

BACKGROUND: We tested the repeatability of myocardial blood flow (MBF) quantified using 82Rb with and without motion correction (MC) and with arterial input functions estimated from left ventricle (LV) and atrium (LA).

METHODS: Twenty-one patients referred for clinical 82Rb PET/CT underwent repeated rest scans in a single imaging session. Global MBF was quantified using three different assessments by two operators: (1) automatic processing without MC and LV arterial input function (AIF), (2) with MC and LV-AIF, and (3) with MC and LA-AIF. Inter-scan and inter-operator repeatability were tested using coefficient of variation (CV).

RESULTS: MC with LV-AIF did not change MBF (no MC: 1.01 ± 0.30 mL/min/g vs MC with LV-AIF: 1.01 ± 0.29, P = 0.70), whereas MC with LA-AIF showed significantly lower MBF assessments (0.95 ± 0.28 mL/min/g, P = 0.0006). We report significant improvement for test-retest reproducibility for global MBF following MC (CV; No MC: 16.0, MC (LV-AIF): 9.2, MC (LA-AIF): 8.8). Good inter-operator repeatability was observed for LV-AIF (CV = 4.7) and LA-AIF (CV = 5.6) for global MBF assessments.

CONCLUSIONS: MC significantly improved the test-retest repeatability between operators and between scans. MBF obtained after MC with LV-AIF were comparable, whereas MBFs after MC and LA-AIF were significantly reduced.

Original languageEnglish
JournalJournal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology
Volume28
Issue number4
Pages (from-to)1718-1725
Number of pages8
ISSN1071-3581
DOIs
Publication statusPublished - Aug 2021

Keywords

  • Myocardial blood flow
  • arterial input function
  • positron emission tomography
  • rubidium

Fingerprint

Dive into the research topics of 'Short-term repeatability of myocardial blood flow using 82Rb PET/CT: The effect of arterial input function position and motion correction'. Together they form a unique fingerprint.

Cite this