Research
Print page Print page
Switch language
The Capital Region of Denmark - a part of Copenhagen University Hospital
Published

Sensitivity analysis of magnetic field measurements for magnetic resonance electrical impedance tomography (MREIT)

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Three-dimensional accelerated acquisition for hyperpolarized 13 C MR  with blipped stack-of-spirals and conjugate-gradient  SENSE

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. ActiveAxADD: Toward non-parametric and orientationally invariant axon diameter distribution mapping using PGSE

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Quantification of cerebral perfusion and cerebrovascular reserve using Turbo-QUASAR arterial spin labeling MRI

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Coil profile estimation strategies for parallel imaging with hyperpolarized 13 C MRI

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Accurate and robust whole-head segmentation from magnetic resonance images for individualized head modeling

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Stimulating aged brains with transcranial direct current stimulation: Opportunities and challenges

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Multi-site benchmarking of clinical 13C RF coils at 3T

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Three-dimensional accelerated acquisition for hyperpolarized 13 C MR  with blipped stack-of-spirals and conjugate-gradient  SENSE

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. Guidelines for TMS/tES Clinical Services and Research through the COVID-19 Pandemic

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

PURPOSE: Clinical use of magnetic resonance electrical impedance tomography (MREIT) still requires significant sensitivity improvements. Here, the measurement of the current-induced magnetic field (ΔBz,c ) is improved using systematic efficiency analyses and optimization of multi-echo spin echo (MESE) and steady-state free precession free induction decay (SSFP-FID) sequences.

THEORY AND METHODS: Considering T1 , T2 , and T2* relaxation in the signal-to-noise ratios (SNRs) of the MR magnitude images, the efficiency of MESE and SSFP-FID MREIT experiments, and its dependence on the sequence parameters, are analytically analyzed and simulated. The theoretical results are experimentally validated in a saline-filled homogenous spherical phantom with relaxation parameters similar to brain tissue. Measurement of ΔBz,c is also performed in a cylindrical phantom with saline and chicken meat.

RESULTS: The efficiency simulations and experimental results are in good agreement. When using optimal parameters, ΔBz,c can be reliably measured in the phantom even at injected current strengths of 1 mA or lower for both sequence types. The importance of using proper crusher gradient selection on the phase evolution in a MESE experiment is also demonstrated.

CONCLUSION: The efficiencies observed with the optimized sequence parameters will likely render in-vivo human brain MREIT feasible. Magn Reson Med, 2017. © 2017 International Society for Magnetic Resonance in Medicine.