Skip to main navigation Skip to search Skip to main content

Redo Transcatheter Aortic Valve Implantation with the ALLEGRA Transcatheter Heart Valve: Insights from Bench Testing

Mariama Akodad, Maximilian Kütting, Stephanie Sellers, Alina Kirsten, Philipp Marx, Isabel Kim, Anson Cheung, Jonathon Leipsic, Lars Søndergaard, Stefan Toggweiler, David A Wood, John G Webb, Janarthanan Sathananthan

1 Citation (Scopus)

Abstract

PURPOSE: Failure of transcatheter heart valves (THV) may potentially be treated with repeat transcatheter aortic valve implantation (redo TAVI). We assessed hydrodynamic performance, stability and pinwheeling utilizing the ALLEGRA (New Valve Technology, Hechingen, Germany) THV, a CE approved and marketed THV in Europe, inside different THVs.

METHODS: Redo TAVI was simulated with the 27 mm ALLEGRA THV at three implantation depths (-4 mm, 0 mm and +4 mm) in seven different 'failed' THVs: 26 mm Evolut Pro, 25 mm Lotus, 25 mm JenaValve, 25 mm Portico, 23 mm Sapien 3, 27 mm ALLEGRA and M ACURATE neo. Hydrodynamic evaluation was performed according to International Standards Organization 5840-3:2021.

RESULTS: The ALLEGRA THV was stable with acceptable performance (gradient <20 mmHg, effective orifice area >2 cm2, and regurgitant fraction <20%) in all 'failed' THVs except the Evolut Pro at -4 mm implantation depth. In this configuration, the outflow of the ALLEGRA frame was constrained by the Evolut Pro THV and the ALLEGRA leaflets were unable to fully close. Pinwheeling was severe for the ALLEGRA in Evolut Pro. The neo-skirt was higher with taller frame THVs.

CONCLUSION: The ALLEGRA THV had favorable hydrodynamic performance, stability and pinwheeling in all redo TAVI samples except the Evolut Pro at low implantation depth with compromised function. The choice of initial THV may have late implications on new THV choice and function.

Original languageEnglish
JournalCardiovascular engineering and technology
Volume13
Issue number6
Pages (from-to)930-938
Number of pages9
ISSN1869-408X
DOIs
Publication statusPublished - Dec 2022

Keywords

  • Aortic Valve Stenosis
  • Aortic Valve/diagnostic imaging
  • Heart Valve Prosthesis
  • Humans
  • Prosthesis Design
  • Transcatheter Aortic Valve Replacement/adverse effects
  • Treatment Outcome

Fingerprint

Dive into the research topics of 'Redo Transcatheter Aortic Valve Implantation with the ALLEGRA Transcatheter Heart Valve: Insights from Bench Testing'. Together they form a unique fingerprint.

Cite this