Quantification of the sensitivity increase of a micro-high-performance liquid chromatography-electrospray ionization mass spectrometry system with decreasing column diameter

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Separation of steroid isomers by ion mobility mass spectrometry

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Impact of binge drinking on hepatic lipid metabolism: lipidome analysis of liver vein and peripheral blood during acute alcohol intoxication

    Research output: Contribution to journalConference abstract in journalResearchpeer-review

  2. Polyols and Branched Chained Amino Acids Are Associated with Present and Future Renal Impairment in Type 1 Diabetes

    Research output: Contribution to journalConference abstract in journalResearchpeer-review

  3. High sensitivity LC-MS profiling of antibody-drug conjugates with difluoroacetic acid ion pairing

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Cell populations and molecular biomarkers in blood and urine characterise nephropathy in type 1 diabetes

    Research output: Contribution to journalConference abstract in journalResearchpeer-review

  5. Circulating metabolites are associated with present and future renal impairment in type 1 Diabetes

    Research output: Contribution to journalConference abstract in journalResearchpeer-review

View graph of relations

This study details the sensitivity achieved with capillary columns when used with a micro-HPLC-electrospray ionization MS system. It is comprised of two sections, the first is the comparative study of three columns, one of narrow-bore diameter and two of capillary diameter. The second section compares three columns of decreasing diameter in the capillary scale. All the experiments achieved enhanced sensitivity using capillary columns. The increase in the experimental MS response ranged from -20% to +20% compared to the UV experimental response when decreasing the internal diameter of the columns used. When comparing the experimental MS response to the maximum theoretical UV response achievable, the increase in response ranged from 40 to 50%.

Original languageEnglish
JournalJournal of Chromatography
Volume976
Issue number1-2
Pages (from-to)11-8
Number of pages8
ISSN0021-9673