Pulse contour-derived cardiac output in hemodialysis patients

Joakim Cordtz, Soeren D Ladefoged

    5 Citations (Scopus)

    Abstract

    Reliable methods for cardiac output determination are essential for studying the pathophysiology of intradialytic hypotension. Use of the current gold standard, the Transonic monitor, requires an arteriovenous fistula. We wished to verify the accuracy of a method based on finger pulse contour analysis, namely the Finometer monitor (FNM) for further use on patients dialyzing on a central vascular catheter. Fifty simultaneous cardiac output measurements were obtained during hemodialysis sessions in 25 patients. The internal variability of the FNM measurements was assessed by comparing 24 pairs of immediately successive measurements. The variability of successive FNM measurements was small (bias 0.28%, SD +/- 6.1%; NS). The absolute cardiac output values reported by the FNM were unreliable (bias 20.1%, SD +/- 35.3%; P
    Original languageEnglish
    JournalHemodialysis International
    Volume14
    Issue number1
    Pages (from-to)78-83
    Number of pages6
    ISSN1492-7535
    DOIs
    Publication statusPublished - 1 Jan 2010