Portable Vector Flow Imaging Compared with Spectral Doppler Ultrasonography

Research output: Contribution to journalJournal articleResearchpeer-review

  1. In vivo Motion Correction in Super Resolution Imaging of Rat Kidneys

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Automatic Detection of B-lines in In-Vivo Lung Ultrasound

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Real-Time 2-D Phased Array Vector Flow Imaging

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Noninvasive Estimation of Pressure Changes Using 2-D Vector Velocity Ultrasound: An Experimental Study With In Vivo Examples

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. A Vector Flow Imaging Method for Portable Ultrasound Using Synthetic Aperture Sequential Beamforming

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Pressure Difference Estimation in Non-stenotic Carotid Bifurcation Phantoms Using Vector Flow Imaging

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Evaluation of 2D super-resolution ultrasound imaging of the rat renal vasculature using ex vivo micro-computed tomography

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Highlights of the development in ultrasound during the last 70 years: A historical review

    Research output: Contribution to journalReviewResearchpeer-review

View graph of relations

In this study, a vector flow imaging (VFI) method developed for a portable ultrasound scanner was used for estimating peak velocity values and variation in beam-to-flow angle over the cardiac cycle in vivo on healthy volunteers. Peak-systolic velocity (PSV), end-diastolic velocity (EDV), and resistive index (RI) measured with VFI were compared to spectral Doppler ultrasonography (SDU). Seventeen healthy volunteers were scanned on the left and right common carotid arteries (CCAs). The standard deviation (SD) of VFI measurements averaged over the cardiac cycle was 7.3% for the magnitude and 3.84° for the angle. Bland-Altman plots showed a positive bias for the PSV measured with SDU (mean difference: 0.31 ms -1 ), and Pearson correlation analysis showed a highly significant correlation ( r = 0.6 ; ). A slightly positive bias was found for EDV and RI measured with SDU (mean difference: 0.08 ms -1 and -0.01 ms -1 , respectively). However, the correlation was low and not significant. The beam-to-flow angle was estimated over the systolic part of the cardiac cycle, and its variations were for all measurements larger than the precision of the angle estimation. The range spanned deviations from -25.2° (-6.0 SD) to 23.7° (4.2 SD) with an average deviation from -15.2° to 9.7°. This can significantly affect PSV values measured by SDU as the beam-to-flow angle is not constant and not aligned with the vessel surface. The study demonstrates that the proposed VFI method can be used in vivo for the measurement of PSV in the CCAs, and that angle variations across the cardiac cycle can lead to significant errors in SDU velocity estimates.

Original languageEnglish
Article number8478145
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Volume66
Issue number3
Pages (from-to)453-462
Number of pages10
ISSN0885-3010
DOIs
Publication statusPublished - Mar 2019