T1 - Performance of a dual-hormone closed-loop system versus insulin-only closed-loop system in adolescents with type 1 diabetes. A single-blind, randomized, controlled, crossover trial
AU - Lindkvist, Emilie Bundgaard
AU - Laugesen, Christian
AU - Reenberg, Asbjørn Thode
AU - Ritschel, Tobias Kasper Skov
AU - Svensson, Jannet
AU - Jørgensen, John Bagterp
AU - Nørgaard, Kirsten
AU - Ranjan, Ajenthen G.
PY - 2023/1/23
Y1 - 2023/1/23
N2 - ObjectiveTo assess the efficacy and safety of a dual-hormone (DH [insulin and glucagon]) closed-loop system compared to a single-hormone (SH [insulin only]) closed-loop system in adolescents with type 1 diabetes.MethodsThis was a 26-hour, two-period, randomized, crossover, inpatient study involving 11 adolescents with type 1 diabetes (nine males [82%], mean ± SD age 14.8 ± 1.4 years, diabetes duration 5.7 ± 2.3 years). Except for the treatment configuration of the DiaCon Artificial Pancreas: DH or SH, experimental visits were identical consisting of: an overnight stay (10:00 pm until 7:30 am), several meals/snacks, and a 45-minute bout of moderate intensity continuous exercise. The primary endpoint was percentage of time spent with sensor glucose values below range (TBR [ResultsOverall, there were no differences between DH and SH for the following glycemic outcomes (median [IQR]): TBR 1.6 [0.0, 2.4] vs. 1.28 [0.16, 3.19]%, p=1.00; time in range (TIR [3.9-10.0 mmol/L]) 68.4 [48.7, 76.8] vs. 75.7 [69.8, 87.1]%, p=0.08; and time above range (TAR [>10.0 mmol/L]) 28.1 [18.1, 49.8] vs. 23.3 [12.3, 27.2]%, p=0.10. Mean ( ± SD) glucose was higher during DH than SH (8.7 ( ± 3.2) vs. 8.1 ( ± 3.0) mmol/L, pConclusionsDH and SH performed similarly in adolescents with type 1 diabetes during a 26-hour inpatient monitoring period involving several metabolic challenges including feeding and exercise. However, during the night and around exercise, the SH configuration outperformed DH.
AB - ObjectiveTo assess the efficacy and safety of a dual-hormone (DH [insulin and glucagon]) closed-loop system compared to a single-hormone (SH [insulin only]) closed-loop system in adolescents with type 1 diabetes.MethodsThis was a 26-hour, two-period, randomized, crossover, inpatient study involving 11 adolescents with type 1 diabetes (nine males [82%], mean ± SD age 14.8 ± 1.4 years, diabetes duration 5.7 ± 2.3 years). Except for the treatment configuration of the DiaCon Artificial Pancreas: DH or SH, experimental visits were identical consisting of: an overnight stay (10:00 pm until 7:30 am), several meals/snacks, and a 45-minute bout of moderate intensity continuous exercise. The primary endpoint was percentage of time spent with sensor glucose values below range (TBR [ResultsOverall, there were no differences between DH and SH for the following glycemic outcomes (median [IQR]): TBR 1.6 [0.0, 2.4] vs. 1.28 [0.16, 3.19]%, p=1.00; time in range (TIR [3.9-10.0 mmol/L]) 68.4 [48.7, 76.8] vs. 75.7 [69.8, 87.1]%, p=0.08; and time above range (TAR [>10.0 mmol/L]) 28.1 [18.1, 49.8] vs. 23.3 [12.3, 27.2]%, p=0.10. Mean ( ± SD) glucose was higher during DH than SH (8.7 ( ± 3.2) vs. 8.1 ( ± 3.0) mmol/L, pConclusionsDH and SH performed similarly in adolescents with type 1 diabetes during a 26-hour inpatient monitoring period involving several metabolic challenges including feeding and exercise. However, during the night and around exercise, the SH configuration outperformed DH.
KW - Adolescent
KW - Blood Glucose/metabolism
KW - Cross-Over Studies
KW - Diabetes Mellitus, Type 1/drug therapy
KW - Glucose
KW - Humans
KW - Insulin
KW - Male
KW - Single-Blind Method
KW - moderate intensity continuous exercise
KW - type 1 diabetes mellitus
KW - adolescents
KW - advanced hybrid closed-loop
KW - non-linear model predictive control
KW - dual-hormone
KW - artificial pancreas