Abstract
BACKGROUND: Left ventricular ejection fraction (LVEF) estimation using adenosine stress myocardial perfusion imaging (MPI) can be challenging. The short half-life of adenosine and the guideline-recommended adenosine infusion stop during Rubidium-82 acquisition protocol may affect the accuracy and repeatability of the LVEF measures.
METHODS: This study comprised 25 healthy volunteers (median age 23 years) who underwent repeat myocardial perfusion imaging (MPI) sessions employing Rubidium-82 PET/CT. A guideline-recommended reconstruction protocol was used for both rest and adenosine stress MPI (150-360 s post-radiotracer injection, standardrecon). For the stress MPI protocol, two additional reconstruction protocols were considered; one was employing 60 seconds data (150-210 seconds, shortfixed) and the other a dynamic frame window based on the bolus arrival of Rubidium-82 in the heart until 210 seconds (x-210 seconds, shortindividual). We report rest and stress LVEF, the LVEF reserve, and the LVEF reserve repeatability.
RESULTS: Differences in the LVEF assessments were observed between the guideline recommended and alternative reconstruction protocol (LVEF stress MPI: standardrecon = 68 ± 7%, shortfixed = 71 ± 7% (P = .08), shortindividual = 72 ± 7% (P = .04)), and the LVEF reserve was reduced for the guideline-recommended protocol (standardrecon = 7.8 ± 3.5, shortfixed = 10.1 ± 3.7, shortindividual = 10.5 ± 3.6, all P < .001). The best repeatability measures were obtained for the shortindividual protocol (repeatability: standardrecon = 45.3%, shortfixed = 41.2%, shortindividual = 31.7%).
CONCLUSION: We recommend using the shortindividual reconstruction protocol for improved LVEF repeatability and reserve assessment. Alternatively, in centers with limited technical support we recommend the use of the shortfixed protocol.
| Original language | English |
|---|---|
| Journal | Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology |
| Volume | 29 |
| Issue number | 6 |
| Pages (from-to) | 3369-3378 |
| Number of pages | 10 |
| ISSN | 1071-3581 |
| DOIs | |
| Publication status | Published - Dec 2022 |
Keywords
- Adenosine
- Adult
- Heart Ventricles/diagnostic imaging
- Humans
- Myocardial Perfusion Imaging/methods
- Positron Emission Tomography Computed Tomography
- Positron-Emission Tomography/methods
- Rubidium
- Stroke Volume
- Ventricular Function, Left
- Young Adult