TY - JOUR
T1 - Online adaptive radiotherapy of urinary bladder cancer with full re-optimization to the anatomy of the day
T2 - Initial experience and dosimetric benefits
AU - Åström, Lina M
AU - Behrens, Claus P
AU - Calmels, Lucie
AU - Sjöström, David
AU - Geertsen, Poul
AU - Mouritsen, Lene Sonne
AU - Serup-Hansen, Eva
AU - Lindberg, Henriette
AU - Sibolt, Patrik
N1 - Copyright © 2022 The Author(s). Published by Elsevier B.V. All rights reserved.
PY - 2022/6
Y1 - 2022/6
N2 - BACKGROUND AND PURPOSE: Online adaptive radiotherapy (oART) potentially reduces the dose to organs at risk (OARs) as the planning target volume (PTV) margins are reduced compared to a non-adaptive approach (non-ART). This study evaluates the feasibility and dosimetric impact of cone-beam computed tomography (CBCT)-guided oART of urinary bladder cancer for the first patients treated, using patient-specific margins.MATERIALS AND METHODS: Sixteen consecutive patients with muscle-invasive bladder cancer received two or more (median = 23) fractions as oART, and remaining fractions as non-ART. The non-ART fractions were delivered with standard population-based margins, while reduced patient-specific margins based on intra-fractional variations extracted from 2-4 fractions were applied to the primary PTV (PTV-T) during the oART fractions. Target volume and coverage, and dose to OARs were compared between non-ART and oART plans, and the oART procedure time was recorded.RESULTS: In total, 297/512 fractions were delivered as oART with full re-optimization to the anatomy of the day. The median (interquartile range, IQR) oART procedure time, measured from the end of CBCT generation to completion of plan review, and quality assurance was 13.9 (11.9;16.6) min. The median (IQR) volume reduction in PTV-T volume was 33.9 (24.2;45.0)%, comparing oART and non-ART plans, resulting in median (IQR) reductions in bowel bag V45Gy of 18.8 (12.7;27.9)% and rectum V50Gy of 70.7 (35.9;94.8)%. By re-optimizing the plan to the daily anatomy, full target coverage was achieved at all oART fractions.CONCLUSIONS: oART resulted in large reductions in treatment volumes and doses to OARs, compared to non-ART, while ensuring target coverage. This indicates potential reductions in gastrointestinal toxicity.
AB - BACKGROUND AND PURPOSE: Online adaptive radiotherapy (oART) potentially reduces the dose to organs at risk (OARs) as the planning target volume (PTV) margins are reduced compared to a non-adaptive approach (non-ART). This study evaluates the feasibility and dosimetric impact of cone-beam computed tomography (CBCT)-guided oART of urinary bladder cancer for the first patients treated, using patient-specific margins.MATERIALS AND METHODS: Sixteen consecutive patients with muscle-invasive bladder cancer received two or more (median = 23) fractions as oART, and remaining fractions as non-ART. The non-ART fractions were delivered with standard population-based margins, while reduced patient-specific margins based on intra-fractional variations extracted from 2-4 fractions were applied to the primary PTV (PTV-T) during the oART fractions. Target volume and coverage, and dose to OARs were compared between non-ART and oART plans, and the oART procedure time was recorded.RESULTS: In total, 297/512 fractions were delivered as oART with full re-optimization to the anatomy of the day. The median (interquartile range, IQR) oART procedure time, measured from the end of CBCT generation to completion of plan review, and quality assurance was 13.9 (11.9;16.6) min. The median (IQR) volume reduction in PTV-T volume was 33.9 (24.2;45.0)%, comparing oART and non-ART plans, resulting in median (IQR) reductions in bowel bag V45Gy of 18.8 (12.7;27.9)% and rectum V50Gy of 70.7 (35.9;94.8)%. By re-optimizing the plan to the daily anatomy, full target coverage was achieved at all oART fractions.CONCLUSIONS: oART resulted in large reductions in treatment volumes and doses to OARs, compared to non-ART, while ensuring target coverage. This indicates potential reductions in gastrointestinal toxicity.
KW - Cone-Beam Computed Tomography/methods
KW - Humans
KW - Organs at Risk
KW - Radiotherapy Dosage
KW - Radiotherapy Planning, Computer-Assisted/methods
KW - Radiotherapy, Image-Guided/methods
KW - Radiotherapy, Intensity-Modulated/methods
KW - Urinary Bladder Neoplasms/radiotherapy
U2 - 10.1016/j.radonc.2022.03.014
DO - 10.1016/j.radonc.2022.03.014
M3 - Journal article
C2 - 35358605
SN - 0167-8140
VL - 171
SP - 37
EP - 42
JO - Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
JF - Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
ER -