Influence of sample centrifugation on plasma platelet count and activated partial thromboplastin time using patient samples

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Uncarboxylated Matrix Gla-Protein: A Biomarker of Vitamin K Status and Cardiovascular Risk

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Abnormal routine blood tests as predictors of mortality in acutely admitted patients

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Order of draw practices in venous blood sampling at clinical biochemistry departments in the Danish health care system

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Preliminary report: Measurement of apixaban and rivaroxaban in plasma from bleeding patients

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

BACKGROUND: Diagnostic coagulation testing is vulnerable to factors of the pre-analytical phase such as sample centrifugation. Despite this, centrifugation conditions differ widely among European laboratories. Here we use samples from patients referred for Activated partial thromboplastin time (APTT) testing to investigate if different centrifugation conditions result in platelet-poor plasma (PPP) (plasma platelet count < 10 × 109/L) and how the variation in centrifugation conditions affect APTT measurements.

METHODS: Centrifugation of 2000g (10 min) were compared with 3000g (10 min) using samples from patients referred for APTT testing (n = 70). Plasma platelet count and APTT were measured to investigate the influence of the centrifugation conditions. Differences were evaluated using Bland Altman Plots and Student's t-test.

RESULTS: Centrifugation at 3000g for 10 min produced PPP for more of the samples (64%) than centrifugation at 2000g (6%) (p < 0.001). No statistically significant difference for APTT (p = 0.265) was found for samples with APTT < 37 s while samples with prolonged APTT (>37 s) showed a statistically significant difference (p = 0.025). The Bland Altman plot did not reveal a clinically significant difference (mean difference 0.30 s/0.68%) when compared to a maximum acceptable bias of 10%.

CONCLUSION: None of the centrifugation conditions used in this study adequately secured PPP for all samples. Despite a statistically significant difference between samples with prolonged APTT, no clinically significant difference was observed when comparing all APTT measurements.

Original languageEnglish
JournalClinical Biochemistry
Volume83
Pages (from-to)74-77
Number of pages4
ISSN0009-9120
DOIs
Publication statusPublished - Sep 2020