Inductive measurement and encoding of k-space trajectories in MR raw data

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Survey of 800+ data sets from human tissue and body fluid reveals xenomiRs are likely artifacts

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Coil profile estimation strategies for parallel imaging with hyperpolarized 13 C MRI

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. The stray magnetic fields in Magnetic Resonance Current Density Imaging (MRCDI)

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. MR-skanning til diagnostik af nonalkoholisk fedtleversygdom

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Gamma-aminobutyric acid edited echo-planar spectroscopic imaging (EPSI) with MEGA-sLASER at 7T

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

OBJECTIVES: The objective of this study was to concurrently acquire an inductive k-space trajectory measure and corresponding imaging data by an MR scanner.

MATERIALS AND METHODS: 1D gradient measures were obtained by digital integration, regularized using measured gradient coil currents and recorded individually by the scanner concurrently with raw MR data. Gradient measures were frequency modulated into an RF signal receivable by the scanner, yielding a k-space trajectory measure from the cumulative phase of the acquired data. Generation of the gradient measure and frequency modulation was performed by previously developed custom, versatile circuitry.

RESULTS: For a normal echo planar imaging (EPI) sequence, the acquired k-space trajectory measure yielded slightly improved image quality compared to that obtained from using the scanner's estimated eddy current-compensated k-space trajectory. For a spiral trajectory, the regularized inductive k-space trajectory measure lead to a 76% decrease in the root-mean-square error of the reconstructed image.

DISCUSSION: While the proof-of-concept experiments show potential for further improvement, the feasibility of inductively measuring k-space trajectories and increasing the precision through regularization was demonstrated. The approach may offer an inexpensive method to acquire k-space trajectories concurrently with scanning.

Original languageEnglish
JournalMagma (New York, N.Y.)
Volume32
Issue number6
Pages (from-to)655-667
Number of pages13
ISSN0968-5243
DOIs
Publication statusPublished - Dec 2019

    Research areas

  • Encoding of signals in MRI raw data, Gradient imperfections, Inductive measurement of k-space trajectories, Magnetic resonance imaging, Regularization by current measure