Indirect Calorimetry in Mechanically Ventilated Patients: A Prospective, Randomized, Clinical Validation of 2 Devices Against a Gold Standard

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Outcomes of Prophylactic Pantoprazole in Adult Intensive Care Unit Patients Receiving Dialysis: Results of a Randomized Trial

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Time to onset of gastrointestinal bleeding in the SUP-ICU trial: a preplanned substudy

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Clinical Examination for the Prediction of Mortality in the Critically Ill: The Simple Intensive Care Studies-I

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Heterogeneity of treatment effect of stress ulcer prophylaxis in ICU patients: A secondary analysis protocol

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

BACKGROUND AND AIMS: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable.

METHODS: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ([Formula: see text]O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device.

RESULTS: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31% overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142-627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7% overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement -249 to 404 kcal/d.

CONCLUSIONS: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of [Formula: see text]O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.

Original languageEnglish
JournalJournal of Parenteral and Enteral Nutrition
Volume41
Issue number8
Pages (from-to)1272-1277
ISSN0148-6071
DOIs
Publication statusPublished - 2017