In vivo evaluation of the reproducibility of T1 and T2 measured in the brain of patients with multiple sclerosis

17 Citations (Scopus)

Abstract

The precision (reproducibility) of relaxation times derived from magnetic resonance images of patients with multiple sclerosis (MS) were investigated. Measurements of 10 MS patients were performed at 1.5 T on two occasions within 1 wk. T1 and T2 was measured using a partial saturation inversion recovery sequence (6 points) and a Carr-Purcell-Meiboom-Gill phase alternating-phase shift multiple spin-echo sequence with 32 echoes. Regions of interest (ROI) were placed both in apparently normal white matter and plaques. The precision (+/- 1.96 SD) and the confidence intervals for T1 and T2 for white matter and plaques were calculated. The precision of T1 for white matter and plaques was respectively +/- 94 msec and +/- 208 msec. The precision of T2 for white matter and plaques was respectively +/- 18 msec and +/- 26 msec. For all measurements the coefficient of variation was about 9%. Judging from our own study and others as well, a precision better than 10% for T1 and T2 would seem unrealistic at present.

Original languageEnglish
JournalMagnetic Resonance Imaging
Volume10
Issue number4
Pages (from-to)579-84
Number of pages6
ISSN0730-725X
DOIs
Publication statusPublished - 1992
Externally publishedYes

Keywords

  • Adult
  • Brain/pathology
  • Humans
  • Magnetic Resonance Imaging
  • Middle Aged
  • Multiple Sclerosis/diagnosis
  • Reproducibility of Results

Fingerprint

Dive into the research topics of 'In vivo evaluation of the reproducibility of T1 and T2 measured in the brain of patients with multiple sclerosis'. Together they form a unique fingerprint.

Cite this