Research
Print page Print page
Switch language
The Capital Region of Denmark - a part of Copenhagen University Hospital
Published

Improvement in diagnostic quality of structural and angiographic MRI of the brain using motion correction with interleaved, volumetric navigators

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Novel functions of the luteinizing hormone/chorionic gonadotropin receptor in prostate cancer cells and patients

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Inflammation, non-endothelial dependent coronary microvascular function and diastolic function-Are they linked?

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Lipidomics of human adipose tissue reveals diversity between body areas

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Cerebral Oxygen Metabolism in Adults with Sickle Cell Disease

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Quantification of cerebral perfusion and cerebrovascular reserve using Turbo-QUASAR arterial spin labeling MRI

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. European Ultrahigh-Field Imaging Network for Neurodegenerative Diseases (EUFIND)

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

INTRODUCTION: Subject movements lead to severe artifacts in magnetic resonance (MR) brain imaging. In this study we evaluate the diagnostic image quality in T1-weighted, T2-weighted, and time-of-flight angiographic MR sequences when using a flexible, navigator-based prospective motion correction system (iMOCO).

METHODS: Five healthy volunteers were scanned during different movement scenarios with and without (+/-) iMOCO activated. An experienced neuroradiologist graded images for image quality criteria (grey-white-matter discrimination, basal ganglia, and small structure and vessel delineation), and general image quality on a four-grade scale.

RESULTS: In scans with deliberate motion, there was a significant improvement in the image quality with iMOCO compared to the scans without iMOCO in both general image impression (T1 p<0.01, T2 p<0.01, TOF p = 0.03) and in anatomical grading (T1 p<0.01, T2 p<0.01, TOF p = 0.01). Subjective image quality was considered non-diagnostic in 91% of the scans with motion -iMOCO, but only in 4% of the scans with motion +iMOCO. iMOCO performed best in the T1-weighted sequence and least well in the angiography sequence. iMOCO was not shown to have any negative effect on diagnostic image quality, as no significant difference in diagnostic quality was seen between scans -iMOCO and +iMOCO with no deliberate movement.

CONCLUSION: The evaluation showed that iMOCO enables substantial improvements in image quality in scans affected by subject movement, recovering important diagnostic information in an otherwise unusable scan.

Original languageEnglish
Article numbere0217145
JournalPLoS One
Volume14
Issue number5
Pages (from-to)1-16
Number of pages26
ISSN1932-6203
DOIs
Publication statusPublished - 1 May 2019

ID: 57192638