Optical surface scanning for respiratory motion monitoring in radiotherapy: a feasibility study

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Proceedings of the 2nd BEAT-PCD conference and 3rd PCD training school: part 1

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Automated Hierarchical time gain compensed for in-vivo ultrasound imaging

    Research output: Contribution to journalConference articleResearchpeer-review

  3. Transverse oscillation vector flow imaging for transthoracic echocardiography

    Research output: Contribution to journalConference articleResearchpeer-review

  4. Determining inter-fractional motion of the uterus using 3D ultrasound imaging during radiotherapy for cervical cancer

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Normal and Malignant Cells Exhibit Differential Responses to Calcium Electroporation

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Repeated diffusion MRI reveals earliest time point for stratification of radiotherapy response in brain metastases

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Diffusion-Weighted MRI for Detection of Tissue Electroporation in vivo

    Research output: Chapter in Book/Report/Conference proceedingBook chapterCommunication

  4. Evaluation of uterine ultrasound imaging in cervical radiotherapy; a comparison of autoscan and conventional probe

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations
Purpose. We evaluated the feasibility of a surface scanning system (Catalyst) for respiratory motion monitoring of breast cancer patients treated with radiotherapy in deep inspiration breath-hold (DIBH). DIBH is used to reduce the radiation dose to the heart and lung. In contrast to RPM, a competing marker-based system, Catalyst does not require an objectmarker on the patient’s skin.

Materials and Methods. Experiment 1: a manikin was used to simulate sinusoidal breathing. The amplitude, period and baseline (signal value at end-expiration) were estimated with RPM and Catalyst. Experiment 2 and 3: the Quasar phantom was used to study if the angle of the monitored surface affects the amplitude of the recorded signal.

Results. Experiment 1: we observed comparable period estimates for both systems. The amplitudes were 8 ± 0.1 mm (Catalyst) and 4.9 ± 0.1 mm (RPM). Independent check with in-room lasers showed an amplitude of approximately 8 mm, supporting Catalyst measurements. Large baseline errors were seen with RPM. Experiment 2: RPM underestimated the amplitude if the object-marker was angled during vertical motion. This result explains the amplitude underestimation by RPM seen in Experiment 1. Experiment 3: an increased (fixed) surface angle during breathing motion resulted in an overestimated amplitude with RPM, while the amplitude estimated by Catalyst was unaffected.

Conclusion. Our study showed that Catalyst can be used as a better alternative to the RPM. With Catalyst, the amplitude estimates are more accurate and do not depend on the angle of the tracked surface, and the baseline errors are smaller.
Original languageEnglish
Article number90360I
JournalSpie Proceedings
Volume9036
DOIs
Publication statusPublished - 2014

ID: 45458423