Effect of Attenuation Correction on Regional Quantification Between PET/MR and PET/CT: A Multicenter Study Using a 3-Dimensional Brain Phantom

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Early Postoperative 18F-FET PET/MRI for Pediatric Brain and Spinal Cord Tumors

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Predictive value of PET response combined with baseline metabolic tumor volume in peripheral T-cell lymphoma patients

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Reproducibility of MR-based Attenuation Maps in PET/MRI and the Impact on PET Quantification in Lung Cancer

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Feasibility of multi-parametric PET and MRI for prediction of tumour recurrence in patients with glioblastoma

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Very Early Response Evaluation by PET/MR in Patients with Lung Cancer-Timing and Feasibility

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Development of a Symmetric Echo-Planar Spectroscopy Imaging Framework for Hyperpolarized 13C Imaging in a Clinical PET/MR Scanner

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

UNLABELLED: A spatial bias in brain PET/MR exists compared with PET/CT, because of MR-based attenuation correction. We performed an evaluation among 4 institutions, 3 PET/MR systems, and 4 PET/CT systems using an anthropomorphic brain phantom, hypothesizing that the spatial bias would be minimized with CT-based attenuation correction (CTAC).

METHODS: The evaluation protocol was similar to the quantification of changes in neurologic PET studies. Regional analysis was conducted on 8 anatomic volumes of interest (VOIs) in gray matter on count-normalized, resolution-matched, coregistered data. On PET/MR systems, CTAC was applied as the reference method for attenuation correction.

RESULTS: With CTAC, visual and quantitative differences between PET/MR and PET/CT systems were minimized. Intersystem variation between institutions was +3.42% to -3.29% in all VOIs for PET/CT and +2.15% to -4.50% in all VOIs for PET/MR. PET/MR systems differed by +2.34% to -2.21%, +2.04% to -2.08%, and -1.77% to -5.37% when compared with a PET/CT system at each institution, and these differences were not significant (P ≥ 0.05).

CONCLUSION: Visual and quantitative differences between PET/MR and PET/CT systems can be minimized by an accurate and standardized method of attenuation correction. If a method similar to CTAC can be implemented for brain PET/MRI, there is no reason why PET/MR should not perform as well as PET/CT.

Original languageEnglish
JournalJournal of nuclear medicine : official publication, Society of Nuclear Medicine
Volume57
Issue number5
Pages (from-to)818-24
Number of pages7
ISSN0161-5505
DOIs
Publication statusPublished - May 2016