Measuring motion-induced B0 -fluctuations in the brain using field probes

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Comparison of prospective head motion correction with NMR field probes and an optical tracking system

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Gamma-aminobutyric acid edited echo-planar spectroscopic imaging (EPSI) with MEGA-sLASER at 7T

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Improved calculation of the equilibrium magnetization of arterial blood in arterial spin labeling

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Effects of imaging gradients in sequences with varying longitudinal storage time-Case of diffusion exchange imaging

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. Sensitivity analysis of magnetic field measurements for magnetic resonance electrical impedance tomography (MREIT)

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. No trace of phase: Corticomotor excitability is not tuned by phase of pericentral mu-rhythm

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Correction to: Ultra-high field MR angiography in human migraine models: a 3.0 T/7.0 T comparison study

    Research output: Contribution to journalComment/debateResearchpeer-review

  3. Ultra-high field MR angiography in human migraine models: a 3.0 T/7.0 T comparison study

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Functional neuroimaging of recovery from motor conversion disorder: A case report

    Research output: Contribution to journalReviewResearchpeer-review

View graph of relations

PURPOSE: Fluctuations of the background magnetic field (B0 ) due to body and breathing motion can lead to significant artifacts in brain imaging at ultrahigh field. Corrections based on real-time sensing using external field probes show great potential. This study evaluates different aspects of field interpolation from these probes into the brain which is implicit in such methods. Measurements and simulations were performed to quantify how well B0 -fluctuations in the brain due to body and breathing motion are reflected in external field probe measurements.

METHODS: Field probe measurements were compared with scanner acquired B0 -maps from experiments with breathing and shoulder movements. A realistic simulation of B0 -fluctuations caused by breathing was performed, and used for testing different sets of field probe positions.

RESULTS: The B0 -fluctuations were well reflected in the field probe measurements in the shoulder experiments, while the breathing experiments showed only moderate correspondence. The simulations showed the importance of the probe positions, and that performing full 3(rd) order corrections based on 16 field probes is not recommended.

CONCLUSION: Methods for quantitative assessment of the field interpolation problem were developed and demonstrated. Field corrections based on external field measurements show great potential, although potential pitfalls were identified. Magn Reson Med, 2015. © 2015 Wiley Periodicals, Inc.

Original languageEnglish
JournalMagnetic Resonance in Medicine
Volume75
Issue number5
Pages (from-to)2020-2030
Number of pages11
ISSN0740-3194
DOIs
Publication statusPublished - May 2016

ID: 45444817