Research
Print page Print page
Switch language
The Capital Region of Denmark - a part of Copenhagen University Hospital
Published

Evaluation of healthy muscle tissue by strain and shear wave elastography - Dependency on depth and ROI position in relation to underlying bone

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Validation of an assessment tool for estimation of abdominal aortic aneurysm compression in diagnostic ultrasound

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Vector velocity volume flow estimation: Sources of error and corrections applied for arteriovenous fistulas

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. First report on intraoperative vector flow imaging of the heart among patients with healthy and diseased aortic valves

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Internal strain estimation for quantification of human heel pad elastic modulus: A phantom study

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. In-vivo validation of fast spectral velocity estimation techniques

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Pressure Difference Estimation in Non-stenotic Carotid Bifurcation Phantoms Using Vector Flow Imaging

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Evaluation of 2D super-resolution ultrasound imaging of the rat renal vasculature using ex vivo micro-computed tomography

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Highlights of the development in ultrasound during the last 70 years: A historical review

    Research output: Contribution to journalReviewResearchpeer-review

View graph of relations

PURPOSE: The aim of this study was to evaluate the influence of depth and underlying bone on strain ratios and shear wave speeds for three different muscles in healthy volunteers. For strain ratios the influence from different reference region-of-interest positions was also evaluated.

MATERIAL AND METHODS: Ten healthy volunteers (five males and five females) had their biceps brachii, gastrocnemius, and quadriceps muscle examined with strain- and shear wave elastography at three different depths and in regions located above bone and beside bone. Strain ratios were averaged from cine-loops of 10s length, and shear wave speeds were measured 10 times at each target point. The distance from the skin surface to the centre of each region-of-interest was measured. Measurements were evaluated with descriptive statistics and linear regression.

RESULTS: Linear regression showed a significant influence on strain ratio measurements from the reference region-of-interest position, i.e. being above the same structures as the target region-of-interest or not (means: 1.65 and 0.78; (P<0.001)). For shear wave speeds, there was a significant influence from depth and location above or beside bone (P=0.011 and P=0.031).

CONCLUSION: Strain ratio values depend significantly on reference and target region-of-interest being above the same tissue, for instance bone. Strain ratios were not influenced by depth in this study. Shear wave speeds decreased with increasing scanning depth and if there was bone below the region-of-interest.

Original languageEnglish
JournalUltrasonics
Volume71
Pages (from-to)127-33
Number of pages7
ISSN0041-624X
DOIs
Publication statusPublished - Sep 2016

    Research areas

  • Journal Article

ID: 49301602