TY - JOUR
T1 - Efficacy and Safety of Appropriate Shocks and Antitachycardia Pacing in Transvenous and Subcutaneous Implantable Defibrillators
T2 - Analysis of All Appropriate Therapy in the PRAETORIAN Trial
AU - Knops, Reinoud E
AU - van der Stuijt, Willeke
AU - Delnoy, Peter Paul H M
AU - Boersma, Lucas V A
AU - Kuschyk, Juergen
AU - El-Chami, Mikhael F
AU - Bonnemeier, Hendrik
AU - Behr, Elijah R
AU - Brouwer, Tom F
AU - Kaab, Stefan
AU - Mittal, Suneet
AU - Smeding, Lonneke
AU - Tijssen, Jan G P
AU - Bijsterveld, Nick R
AU - Richter, Sergio
AU - Brouwer, Marc A
AU - de Groot, Joris R
AU - Kooiman, Kirsten M
AU - Lambiase, Pier D
AU - Neuzil, Petr
AU - Vernooy, Kevin
AU - Alings, Marco
AU - Betts, Timothy R
AU - Bracke, Frank A L E
AU - Burke, Martin C
AU - de Jong, Jonas S S G
AU - Wright, David J
AU - Jansen, Ward P J
AU - Whinnett, Zachary I
AU - Nordbeck, Peter
AU - Knaut, Michael
AU - Philbert, Berit T
AU - van Opstal, Jurren M
AU - Chicos, Alexandru B
AU - Allaart, Cornelis P
AU - Borger van der Burg, Alida E
AU - Clancy, Jude F
AU - Dizon, Jose M
AU - Miller, Marc A
AU - Nemirovsky, Dmitry
AU - Surber, Ralf
AU - Upadhyay, Gaurav A
AU - Weiss, Raul
AU - de Weger, Anouk
AU - Wilde, Arthur A M
AU - Olde Nordkamp, Louise R A
AU - PRAETORIAN Investigators
PY - 2022/2
Y1 - 2022/2
N2 - BACKGROUND: The PRAETORIAN trial (A Prospective, Randomized Comparison of Subcutaneous and Transvenous Implantable Cardioverter Defibrillator Therapy) showed noninferiority of subcutaneous implantable cardioverter defibrillator (S-ICD) compared with transvenous implantable cardioverter defibrillator (TV-ICD) with regard to inappropriate shocks and complications. In contrast to TV-ICD, S-ICD cannot provide antitachycardia pacing for monomorphic ventricular tachycardia. This prespecified secondary analysis evaluates appropriate therapy and whether antitachycardia pacing reduces the number of appropriate shocks.METHODS: The PRAETORIAN trial was an international, investigator-initiated randomized trial that included patients with an indication for implantable cardioverter defibrillator (ICD) therapy. Patients with previous ventricular tachycardia <170 bpm or refractory recurrent monomorphic ventricular tachycardia were excluded. In 39 centers, 849 patients were randomized to receive an S-ICD (n=426) or TV-ICD (n=423) and were followed for a median of 49.1 months. ICD programming was mandated by protocol. Appropriate ICD therapy was defined as therapy for ventricular arrhythmias. Arrhythmias were classified as discrete episodes and storm episodes (≥3 episodes within 24 hours). Analyses were performed in the modified intention-to-treat population.RESULTS: In the S-ICD group, 86 of 426 patients received appropriate therapy, versus 78 of 423 patients in the TV-ICD group, during a median follow-up of 52 months (48-month Kaplan-Meier estimates 19.4% and 17.5%; P=0.45). In the S-ICD group, 83 patients received at least 1 shock, versus 57 patients in the TV-ICD group (48-month Kaplan-Meier estimates 19.2% and 11.5%; P=0.02). Patients in the S-ICD group had a total of 254 shocks, compared with 228 shocks in the TV-ICD group (P=0.68). First shock efficacy was 93.8% in the S-ICD group and 91.6% in the TV-ICD group (P=0.40). The first antitachycardia pacing attempt successfully terminated 46% of all monomorphic ventricular tachycardias, but accelerated the arrhythmia in 9.4%. Ten patients with S-ICD experienced 13 electrical storms, versus 18 patients with TV-ICD with 19 electrical storms. Patients with appropriate therapy had an almost 2-fold increased relative risk of electrical storms in the TV-ICD group compared with the S-ICD group (P=0.05).CONCLUSIONS: In this trial, no difference was observed in shock efficacy of S-ICD compared with TV-ICD. Although patients in the S-ICD group were more likely to receive an ICD shock, the total number of appropriate shocks was not different between the 2 groups. Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT01296022.
AB - BACKGROUND: The PRAETORIAN trial (A Prospective, Randomized Comparison of Subcutaneous and Transvenous Implantable Cardioverter Defibrillator Therapy) showed noninferiority of subcutaneous implantable cardioverter defibrillator (S-ICD) compared with transvenous implantable cardioverter defibrillator (TV-ICD) with regard to inappropriate shocks and complications. In contrast to TV-ICD, S-ICD cannot provide antitachycardia pacing for monomorphic ventricular tachycardia. This prespecified secondary analysis evaluates appropriate therapy and whether antitachycardia pacing reduces the number of appropriate shocks.METHODS: The PRAETORIAN trial was an international, investigator-initiated randomized trial that included patients with an indication for implantable cardioverter defibrillator (ICD) therapy. Patients with previous ventricular tachycardia <170 bpm or refractory recurrent monomorphic ventricular tachycardia were excluded. In 39 centers, 849 patients were randomized to receive an S-ICD (n=426) or TV-ICD (n=423) and were followed for a median of 49.1 months. ICD programming was mandated by protocol. Appropriate ICD therapy was defined as therapy for ventricular arrhythmias. Arrhythmias were classified as discrete episodes and storm episodes (≥3 episodes within 24 hours). Analyses were performed in the modified intention-to-treat population.RESULTS: In the S-ICD group, 86 of 426 patients received appropriate therapy, versus 78 of 423 patients in the TV-ICD group, during a median follow-up of 52 months (48-month Kaplan-Meier estimates 19.4% and 17.5%; P=0.45). In the S-ICD group, 83 patients received at least 1 shock, versus 57 patients in the TV-ICD group (48-month Kaplan-Meier estimates 19.2% and 11.5%; P=0.02). Patients in the S-ICD group had a total of 254 shocks, compared with 228 shocks in the TV-ICD group (P=0.68). First shock efficacy was 93.8% in the S-ICD group and 91.6% in the TV-ICD group (P=0.40). The first antitachycardia pacing attempt successfully terminated 46% of all monomorphic ventricular tachycardias, but accelerated the arrhythmia in 9.4%. Ten patients with S-ICD experienced 13 electrical storms, versus 18 patients with TV-ICD with 19 electrical storms. Patients with appropriate therapy had an almost 2-fold increased relative risk of electrical storms in the TV-ICD group compared with the S-ICD group (P=0.05).CONCLUSIONS: In this trial, no difference was observed in shock efficacy of S-ICD compared with TV-ICD. Although patients in the S-ICD group were more likely to receive an ICD shock, the total number of appropriate shocks was not different between the 2 groups. Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT01296022.
UR - http://www.scopus.com/inward/record.url?scp=85123968708&partnerID=8YFLogxK
U2 - 10.1161/CIRCULATIONAHA.121.057816
DO - 10.1161/CIRCULATIONAHA.121.057816
M3 - Journal article
C2 - 34779221
VL - 145
SP - 321
EP - 329
JO - Circulation (Baltimore)
JF - Circulation (Baltimore)
SN - 0009-7322
IS - 5
ER -