Skip to main navigation Skip to search Skip to main content

Dynamic bicycle exercise to assess cardiac output at multiple exercise levels during magnetic resonance imaging

Johan Heiberg, Benjamin Asschenfeldt, Marie Maagaard, Steffen Ringgaard

16 Citations (Scopus)

Abstract

PURPOSE: The aim was to establish a method for performing dynamic exercise during magnetic resonance imaging (MRI) using a slowly increasing workload protocol.

METHODS: An ergometer bicycle with a step-wise, exercise protocol was used. Real-time phase-contrast MRI images of the aorta were obtained at each exercise step.

RESULTS: In total, 40 participants completed the exercise protocol to reach a mean maximum cardiac output of 13.7±3.7l/min and a heart rate of 150±16beats/min at the highest exercise level. Less than 1% of scans were discarded due to poor quality.

CONCLUSIONS: Dynamic, high intensity exercise is feasible during MRI.

Original languageEnglish
JournalClinical Imaging
Volume46
Pages (from-to)102-107
Number of pages6
ISSN0730-725X
DOIs
Publication statusPublished - 2017
Externally publishedYes