Diagnostic accuracy of a novel tuberculosis point-of-care urine lipoarabinomannan assay for people living with HIV: A meta-analysis of individual in- and outpatient data

Research output: Contribution to journalJournal articleResearchpeer-review

Harvard

Broger, T, Nicol, MP, Székely, R, Bjerrum, S, Sossen, B, Schutz, C, Opintan, JA, Johansen, IS, Mitarai, S, Chikamatsu, K, Kerkhoff, AD, Macé, A, Ongarello, S, Meintjes, G, Denkinger, CM & Schumacher, SG 2020, 'Diagnostic accuracy of a novel tuberculosis point-of-care urine lipoarabinomannan assay for people living with HIV: A meta-analysis of individual in- and outpatient data', PLOS Medicine, vol. 17, no. 5, pp. e1003113. https://doi.org/10.1371/journal.pmed.1003113

APA

Broger, T., Nicol, M. P., Székely, R., Bjerrum, S., Sossen, B., Schutz, C., Opintan, J. A., Johansen, I. S., Mitarai, S., Chikamatsu, K., Kerkhoff, A. D., Macé, A., Ongarello, S., Meintjes, G., Denkinger, C. M., & Schumacher, S. G. (2020). Diagnostic accuracy of a novel tuberculosis point-of-care urine lipoarabinomannan assay for people living with HIV: A meta-analysis of individual in- and outpatient data. PLOS Medicine, 17(5), e1003113. https://doi.org/10.1371/journal.pmed.1003113

CBE

Broger T, Nicol MP, Székely R, Bjerrum S, Sossen B, Schutz C, Opintan JA, Johansen IS, Mitarai S, Chikamatsu K, Kerkhoff AD, Macé A, Ongarello S, Meintjes G, Denkinger CM, Schumacher SG. 2020. Diagnostic accuracy of a novel tuberculosis point-of-care urine lipoarabinomannan assay for people living with HIV: A meta-analysis of individual in- and outpatient data. PLOS Medicine. 17(5):e1003113. https://doi.org/10.1371/journal.pmed.1003113

MLA

Vancouver

Author

Broger, Tobias ; Nicol, Mark P ; Székely, Rita ; Bjerrum, Stephanie ; Sossen, Bianca ; Schutz, Charlotte ; Opintan, Japheth A ; Johansen, Isik S ; Mitarai, Satoshi ; Chikamatsu, Kinuyo ; Kerkhoff, Andrew D ; Macé, Aurélien ; Ongarello, Stefano ; Meintjes, Graeme ; Denkinger, Claudia M ; Schumacher, Samuel G. / Diagnostic accuracy of a novel tuberculosis point-of-care urine lipoarabinomannan assay for people living with HIV : A meta-analysis of individual in- and outpatient data. In: PLOS Medicine. 2020 ; Vol. 17, No. 5. pp. e1003113.

Bibtex

@article{9d5f68e71bf84f2884117db94aa6ece5,
title = "Diagnostic accuracy of a novel tuberculosis point-of-care urine lipoarabinomannan assay for people living with HIV: A meta-analysis of individual in- and outpatient data",
abstract = "BACKGROUND: Tuberculosis (TB) is the most common cause of death in people living with HIV (PLHIV), yet TB often goes undiagnosed since many patients are not able to produce a sputum specimen, and traditional diagnostics are costly or unavailable. A novel, rapid lateral flow assay, Fujifilm SILVAMP TB LAM (SILVAMP-LAM), detects the presence of TB lipoarabinomannan (LAM) in urine, and is substantially more sensitive for diagnosing TB in PLHIV than an earlier LAM assay (Alere Determine TB LAM lateral flow assay [LF-LAM]). Here, we present an individual participant data meta-analysis of the diagnostic accuracy of SILVAMP-LAM in adult PLHIV, including both published and unpublished data.METHODS AND FINDINGS: Adult PLHIV (≥18 years) were assessed in 5 prospective cohort studies in South Africa (3 cohorts), Vietnam, and Ghana, carried out during 2012 to 2017. Of the 1,595 PLHIV who met eligibility criteria, the majority (61%) were inpatients, median age was 37 years (IQR 30-43), 43% had a CD4 count ≤ 100 cells/μl, and 35% were receiving antiretroviral therapy. Most participants (94%) had a positive WHO symptom screen for TB on enrollment, and 45% were diagnosed with microbiologically confirmed TB, using mycobacterial culture or Xpert MTB/RIF testing of sputum, urine, or blood. Previously published data from inpatients were combined with unpublished data from outpatients. Biobanked urine samples were tested, using blinded double reading, with SILVAMP-LAM and LF-LAM. Applying a microbiological reference standard for assessment of sensitivity, the overall sensitivity for TB detection was 70.7% (95% CI 59.0%-80.8%) for SILVAMP-LAM compared to 34.9% (95% CI 19.5%-50.9%) for LF-LAM. Using a composite reference standard (which included patients with both microbiologically confirmed as well as clinically diagnosed TB), SILVAMP-LAM sensitivity was 65.8% (95% CI 55.9%-74.6%), and that of LF-LAM 31.4% (95% CI 19.1%-43.7%). In patients with CD4 count ≤ 100 cells/μl, SILVAMP-LAM sensitivity was 87.1% (95% CI 79.3%-93.6%), compared to 56.0% (95% CI 43.9%-64.9%) for LF-LAM. In patients with CD4 count 101-200 cells/μl, SILVAMP-LAM sensitivity was 62.7% (95% CI 52.4%-71.9%), compared to 25.3% (95% CI 15.8%-34.9%) for LF-LAM. In those with CD4 count > 200 cells/μl, SILVAMP-LAM sensitivity was 43.9% (95% CI 34.3%-53.9%), compared to 10.9% (95% CI 5.2%-18.4%) for LF-LAM. Using a microbiological reference standard, the specificity of SILVAMP-LAM was 90.9% (95% CI 87.2%-93.7%), and that of LF-LAM 95.3% (95% CI 92.2%-97.7%). Limitations of this study include the use of biobanked, rather than fresh urine samples, and testing by skilled laboratory technicians in research laboratories, rather than at the point of care.CONCLUSIONS: In this study, we found that SILVAMP-LAM identified a substantially higher proportion of TB patients in PLHIV than LF-LAM. The sensitivity of SILVAMP-LAM was highest in patients with CD4 count ≤ 100 cells/μl. Further work is needed to demonstrate accuracy when implemented as a point-of-care test.",
keywords = "Adult, Ambulatory Care Facilities, Female, HIV Infections/complications, Humans, Lipopolysaccharides/analysis, Male, Point-of-Care Systems, Prospective Studies, Tuberculosis/diagnosis, Tuberculosis, Pulmonary/diagnosis",
author = "Tobias Broger and Nicol, {Mark P} and Rita Sz{\'e}kely and Stephanie Bjerrum and Bianca Sossen and Charlotte Schutz and Opintan, {Japheth A} and Johansen, {Isik S} and Satoshi Mitarai and Kinuyo Chikamatsu and Kerkhoff, {Andrew D} and Aur{\'e}lien Mac{\'e} and Stefano Ongarello and Graeme Meintjes and Denkinger, {Claudia M} and Schumacher, {Samuel G}",
year = "2020",
month = may,
doi = "10.1371/journal.pmed.1003113",
language = "English",
volume = "17",
pages = "e1003113",
journal = "PLOS Medicine",
issn = "1549-1277",
publisher = "Public Library of Science",
number = "5",

}

RIS

TY - JOUR

T1 - Diagnostic accuracy of a novel tuberculosis point-of-care urine lipoarabinomannan assay for people living with HIV

T2 - A meta-analysis of individual in- and outpatient data

AU - Broger, Tobias

AU - Nicol, Mark P

AU - Székely, Rita

AU - Bjerrum, Stephanie

AU - Sossen, Bianca

AU - Schutz, Charlotte

AU - Opintan, Japheth A

AU - Johansen, Isik S

AU - Mitarai, Satoshi

AU - Chikamatsu, Kinuyo

AU - Kerkhoff, Andrew D

AU - Macé, Aurélien

AU - Ongarello, Stefano

AU - Meintjes, Graeme

AU - Denkinger, Claudia M

AU - Schumacher, Samuel G

PY - 2020/5

Y1 - 2020/5

N2 - BACKGROUND: Tuberculosis (TB) is the most common cause of death in people living with HIV (PLHIV), yet TB often goes undiagnosed since many patients are not able to produce a sputum specimen, and traditional diagnostics are costly or unavailable. A novel, rapid lateral flow assay, Fujifilm SILVAMP TB LAM (SILVAMP-LAM), detects the presence of TB lipoarabinomannan (LAM) in urine, and is substantially more sensitive for diagnosing TB in PLHIV than an earlier LAM assay (Alere Determine TB LAM lateral flow assay [LF-LAM]). Here, we present an individual participant data meta-analysis of the diagnostic accuracy of SILVAMP-LAM in adult PLHIV, including both published and unpublished data.METHODS AND FINDINGS: Adult PLHIV (≥18 years) were assessed in 5 prospective cohort studies in South Africa (3 cohorts), Vietnam, and Ghana, carried out during 2012 to 2017. Of the 1,595 PLHIV who met eligibility criteria, the majority (61%) were inpatients, median age was 37 years (IQR 30-43), 43% had a CD4 count ≤ 100 cells/μl, and 35% were receiving antiretroviral therapy. Most participants (94%) had a positive WHO symptom screen for TB on enrollment, and 45% were diagnosed with microbiologically confirmed TB, using mycobacterial culture or Xpert MTB/RIF testing of sputum, urine, or blood. Previously published data from inpatients were combined with unpublished data from outpatients. Biobanked urine samples were tested, using blinded double reading, with SILVAMP-LAM and LF-LAM. Applying a microbiological reference standard for assessment of sensitivity, the overall sensitivity for TB detection was 70.7% (95% CI 59.0%-80.8%) for SILVAMP-LAM compared to 34.9% (95% CI 19.5%-50.9%) for LF-LAM. Using a composite reference standard (which included patients with both microbiologically confirmed as well as clinically diagnosed TB), SILVAMP-LAM sensitivity was 65.8% (95% CI 55.9%-74.6%), and that of LF-LAM 31.4% (95% CI 19.1%-43.7%). In patients with CD4 count ≤ 100 cells/μl, SILVAMP-LAM sensitivity was 87.1% (95% CI 79.3%-93.6%), compared to 56.0% (95% CI 43.9%-64.9%) for LF-LAM. In patients with CD4 count 101-200 cells/μl, SILVAMP-LAM sensitivity was 62.7% (95% CI 52.4%-71.9%), compared to 25.3% (95% CI 15.8%-34.9%) for LF-LAM. In those with CD4 count > 200 cells/μl, SILVAMP-LAM sensitivity was 43.9% (95% CI 34.3%-53.9%), compared to 10.9% (95% CI 5.2%-18.4%) for LF-LAM. Using a microbiological reference standard, the specificity of SILVAMP-LAM was 90.9% (95% CI 87.2%-93.7%), and that of LF-LAM 95.3% (95% CI 92.2%-97.7%). Limitations of this study include the use of biobanked, rather than fresh urine samples, and testing by skilled laboratory technicians in research laboratories, rather than at the point of care.CONCLUSIONS: In this study, we found that SILVAMP-LAM identified a substantially higher proportion of TB patients in PLHIV than LF-LAM. The sensitivity of SILVAMP-LAM was highest in patients with CD4 count ≤ 100 cells/μl. Further work is needed to demonstrate accuracy when implemented as a point-of-care test.

AB - BACKGROUND: Tuberculosis (TB) is the most common cause of death in people living with HIV (PLHIV), yet TB often goes undiagnosed since many patients are not able to produce a sputum specimen, and traditional diagnostics are costly or unavailable. A novel, rapid lateral flow assay, Fujifilm SILVAMP TB LAM (SILVAMP-LAM), detects the presence of TB lipoarabinomannan (LAM) in urine, and is substantially more sensitive for diagnosing TB in PLHIV than an earlier LAM assay (Alere Determine TB LAM lateral flow assay [LF-LAM]). Here, we present an individual participant data meta-analysis of the diagnostic accuracy of SILVAMP-LAM in adult PLHIV, including both published and unpublished data.METHODS AND FINDINGS: Adult PLHIV (≥18 years) were assessed in 5 prospective cohort studies in South Africa (3 cohorts), Vietnam, and Ghana, carried out during 2012 to 2017. Of the 1,595 PLHIV who met eligibility criteria, the majority (61%) were inpatients, median age was 37 years (IQR 30-43), 43% had a CD4 count ≤ 100 cells/μl, and 35% were receiving antiretroviral therapy. Most participants (94%) had a positive WHO symptom screen for TB on enrollment, and 45% were diagnosed with microbiologically confirmed TB, using mycobacterial culture or Xpert MTB/RIF testing of sputum, urine, or blood. Previously published data from inpatients were combined with unpublished data from outpatients. Biobanked urine samples were tested, using blinded double reading, with SILVAMP-LAM and LF-LAM. Applying a microbiological reference standard for assessment of sensitivity, the overall sensitivity for TB detection was 70.7% (95% CI 59.0%-80.8%) for SILVAMP-LAM compared to 34.9% (95% CI 19.5%-50.9%) for LF-LAM. Using a composite reference standard (which included patients with both microbiologically confirmed as well as clinically diagnosed TB), SILVAMP-LAM sensitivity was 65.8% (95% CI 55.9%-74.6%), and that of LF-LAM 31.4% (95% CI 19.1%-43.7%). In patients with CD4 count ≤ 100 cells/μl, SILVAMP-LAM sensitivity was 87.1% (95% CI 79.3%-93.6%), compared to 56.0% (95% CI 43.9%-64.9%) for LF-LAM. In patients with CD4 count 101-200 cells/μl, SILVAMP-LAM sensitivity was 62.7% (95% CI 52.4%-71.9%), compared to 25.3% (95% CI 15.8%-34.9%) for LF-LAM. In those with CD4 count > 200 cells/μl, SILVAMP-LAM sensitivity was 43.9% (95% CI 34.3%-53.9%), compared to 10.9% (95% CI 5.2%-18.4%) for LF-LAM. Using a microbiological reference standard, the specificity of SILVAMP-LAM was 90.9% (95% CI 87.2%-93.7%), and that of LF-LAM 95.3% (95% CI 92.2%-97.7%). Limitations of this study include the use of biobanked, rather than fresh urine samples, and testing by skilled laboratory technicians in research laboratories, rather than at the point of care.CONCLUSIONS: In this study, we found that SILVAMP-LAM identified a substantially higher proportion of TB patients in PLHIV than LF-LAM. The sensitivity of SILVAMP-LAM was highest in patients with CD4 count ≤ 100 cells/μl. Further work is needed to demonstrate accuracy when implemented as a point-of-care test.

KW - Adult

KW - Ambulatory Care Facilities

KW - Female

KW - HIV Infections/complications

KW - Humans

KW - Lipopolysaccharides/analysis

KW - Male

KW - Point-of-Care Systems

KW - Prospective Studies

KW - Tuberculosis/diagnosis

KW - Tuberculosis, Pulmonary/diagnosis

U2 - 10.1371/journal.pmed.1003113

DO - 10.1371/journal.pmed.1003113

M3 - Journal article

C2 - 32357197

VL - 17

SP - e1003113

JO - PLOS Medicine

JF - PLOS Medicine

SN - 1549-1277

IS - 5

ER -

ID: 62370466