Comparison of prospective head motion correction with NMR field probes and an optical tracking system

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Quantification of cerebral perfusion and cerebrovascular reserve using Turbo-QUASAR arterial spin labeling MRI

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Coil profile estimation strategies for parallel imaging with hyperpolarized 13 C MRI

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Gamma-aminobutyric acid edited echo-planar spectroscopic imaging (EPSI) with MEGA-sLASER at 7T

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Improved calculation of the equilibrium magnetization of arterial blood in arterial spin labeling

    Research output: Contribution to journalJournal articleResearchpeer-review

  5. Effects of imaging gradients in sequences with varying longitudinal storage time-Case of diffusion exchange imaging

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Value and limitations of intracranial recordings for validating electric field modeling for transcranial brain stimulation

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Accessibility of cortical regions to focal TES: Dependence on spatial position, safety, and practical constraints

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Electric field simulations for transcranial brain stimulation using FEM: an efficient implementation and error analysis

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Blinding is compromised for transcranial direct current stimulation at 1 mA for 20 minutes in young healthy adults

    Research output: Contribution to journalJournal articleResearchpeer-review

  • Martin Eschelbach
  • Ali Aghaeifar
  • Jonas Bause
  • Jonas Handwerker
  • Jens Anders
  • Eva-Maria Engel
  • Axel Thielscher
  • Klaus Scheffler
View graph of relations

PURPOSE: The aim of this study was to compare prospective head motion correction and motion tracking abilities of two tracking systems: Active NMR field probes and a Moiré phase tracking camera system using an optical marker.

METHODS: Both tracking systems were used simultaneously on human subjects. The prospective head motion correction was compared in an MP2RAGE and a gradient echo sequence. In addition, the motion tracking trajectories for three subjects were compared against each other and their correlation and deviations were analyzed.

RESULTS: With both tracking systems motion artifacts were visibly reduced. The precision of the field probe system was on the order of 50 µm for translations and 0.03° for rotations while the camera's was approximately 5 µm and 0.007°. The comparison of the measured trajectories showed close correlation and an average absolute deviation below 500 µm and 0.5°.

CONCLUSION: This study presents the first in vivo comparison between NMR field probes and Moiré phase tracking. For the gradient echo images, the field probes had a similar motion correction performance as the optical tracking system. For the MP2RAGE measurement, however, the camera yielded better results. Still, both tracking systems substantially decreased image artifacts in the presence of subject motion. Thus, the motion tracking modality should be chosen according to the specific requirements of the experiment while considering the desired image resolution, refresh rate, and head coil constraints.

Original languageEnglish
JournalMagnetic Resonance in Medicine
Volume81
Issue number1
Pages (from-to)719-729
Number of pages11
ISSN0740-3194
DOIs
Publication statusPublished - Jan 2019

    Research areas

  • field probes, head motion, Moiré phase tracking, motion correction, prospective, tracking

ID: 54880493