Research
Print page Print page
Switch language
The Capital Region of Denmark - a part of Copenhagen University Hospital
Published

Comparison between temporal and rectal temperature measurement

Research output: Contribution to journalJournal articleResearchpeer-review

  1. Natarbejde og komplikationer i graviditeten

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Discontinuation of antidepressants

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Donorsøgning til Hæmatopoietisk stamcelle transplantation

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Kan prævention udløse depression?

    Research output: Contribution to journalEditorialResearch

  5. Somatiske årsager til akut opståede psykiatriske tilstande hos børn og unge

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. New clinical guidelines on the spinal stabilisation of adult trauma patients - consensus and evidence based

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Die Zervikalstütze für Traumapatienten – seit 2018 obsolet?

    Research output: Contribution to journalLetterResearchpeer-review

  3. Præhospital behandling og transport af patienter

    Research output: Chapter in Book/Report/Conference proceedingBook chapterCommunication

View graph of relations
Introduction – Various digital thermometers for non-invasive use have been used increasingly in Danish hospitals, including the Temporal Artery Thermometer (TAT). However, previous studies have concluded that the accuracy of the TAT is unsatisfying at paediatric, surgical, cancer and intensive care patients. The purpose of this study was to compare the accuracy of the TAT to a conventional rectal thermometer within acutely admitted medical patients at the Emergency Department.
Method - The study was designed as a prospective, comparative research. For two months, 381 patients were included. With no more than 7 minutes apart the temperature was measured first with an Exergen TAT-5000 and then controlled with a rectal thermometer from Omron MC-341-E. The measurements were analysed in a Bland-Altman plot, and the sensitivity and specificity of the TAT were calculated.
Result – The differences between TAT and rectal thermometer spread from -1.7°C – 1.7°C. The mean of the differences was drawn in the Bland-Altman plot through 0.17 with a standard deviation at 0.47. The sensitivity and specificity were calculated to 67 % and 96 %, respectively.
Conclusion – Based on this study, we do not recommend the use of TAT as an alternative to rectal thermometer for non-invasive measuring of the body temperature in acutely admitted medical patients.
Funding and trial registration - All authors received honoraria from The Region Capital of Denmark. Study procedures were approved by the local ethical committee and submitted to www.clinicaltrials.org (NCT01817881).