Application of high resolution synchrotron micro-CT radiation in dental implant osseointegration

  • Camilla Albeck Neldam
  • , Torsten Lauridsen
  • , Alexander Rack
  • , Tore Tranberg Lefolii
  • , Niklas Rye Jørgensen
  • , Robert Feidenhans'l
  • , Else Marie Pinholt
17 Citations (Scopus)

Abstract

The purpose of this study was to describe a refined method using high-resolution synchrotron radiation microtomography (SRmicro-CT) to evaluate osseointegration and peri-implant bone volume fraction after titanium dental implant insertion. SRmicro-CT is considered gold standard evaluating bone microarchitecture. Its high resolution, high contrast, and excellent high signal-to-noise-ratio all contribute to the highest spatial resolutions achievable today. Using SRmicro-CT at a voxel size of 5 μm in an experimental goat mandible model, the peri-implant bone volume fraction was found to quickly increase to 50% as the radial distance from the implant surface increased, and levelled out to approximately 80% at a distance of 400 μm. This method has been successful in depicting the bone and cavities in three dimensions thereby enabling us to give a more precise answer to the fraction of the bone-to-implant contact compared to previous methods.

Original languageEnglish
JournalJournal of Cranio-Maxillofacial Surgery
Volume43
Issue number5
Pages (from-to)682-7
Number of pages6
ISSN1010-5182
DOIs
Publication statusPublished - Jun 2015

Fingerprint

Dive into the research topics of 'Application of high resolution synchrotron micro-CT radiation in dental implant osseointegration'. Together they form a unique fingerprint.