Acquiring beam data for a flattening-filter free linear accelerator using organic scintillators

11 Citations (Scopus)

Abstract

Fibre-coupled organic scintillators have been proven a credible alternative to clinically implemented methods for radiotherapy dosimetry, primarily due to their water equivalence and good spatial resolution. Furthermore, the fast response of the scintillators can be exploited to perform time-resolved dosimetry on a highly detailed level. In this study, we present beam data for a Varian TrueBeam linear accelerator, which is capable of delivering flattening-filter free (FFF1) clinical X-ray beams. The beam data have been acquired using an in-house developed dosimetry system based on fibre-coupled organic scintillators. The presented data exhibit high accuracy and precision when compared with data obtained using commercial dosimetry methods, and agree well with results published in the literature.
Original languageEnglish
JournalRadiation Measurements
Volume56
Issue numberSeptember
Pages (from-to)290-93
Number of pages4
ISSN1350-4487
Publication statusPublished - 2013

Fingerprint

Dive into the research topics of 'Acquiring beam data for a flattening-filter free linear accelerator using organic scintillators'. Together they form a unique fingerprint.