Abdominal aortic aneurysm diameter: a comparison of ultrasound measurements with those from standard and three-dimensional computed tomography reconstruction

Brian J Manning, Thorarinn Kristmundsson, Björn Sonesson, Timothy Resch

77 Citations (Scopus)

Abstract

OBJECTIVE: Aortic aneurysm size is a critical determinant of the need for intervention, yet the maximal diameter will often vary depending on the modality and method of measurement. We aimed to define the relationship between commonly used computed tomography (CT) measurement techniques and those based on current reporting standards and to compare the values obtained with diameter measured using ultrasound (US).

METHODS: CT scans from patients with US-detected aneurysms were analyzed using three-dimensional reconstruction software. Maximal aortic diameter was recorded in the anteroposterior (CT-AP) plane, along the maximal ellipse (CT-ME), perpendicular to the maximal ellipse (CT-PME), or perpendicular to the centerline of flow (CT-PCLF). Diameter measurements were compared with each other and with maximal AP diameter according to US (US-AP). Analysis was performed according to the principles of Bland and Altman. Results are expressed as mean +/- standard deviation.

RESULTS: CT and US scans from 109 patients (92 men, 17 women), with a mean age of 72 +/- 8 years, were included. The mean of each series of readings on CT was significantly larger than the mean US-AP measurement (P < .001), and they also differed significantly from each other (P < .001). The CT-PCLF diameter was larger than CT-AP and CT-PME by mean values of 3.0 +/- 6.6 and 5.9 +/- 6.0 mm, respectively. The CT-ME diameter was larger than CT-PCLF by a mean of 2.4 +/- 5 mm. The US-AP diameter was smaller than CT-AP diameter by 4.2 +/- 4.9 mm, CT-ME by 9.6 +/- 8.0 mm, CT-PME by 1.3 +/- 5 mm, and smaller than CT-PCLF by 7.3 +/- 7.0 mm. Aneurysm size did not significantly affect these differences. Seventy-eight percent of 120 pairs of intraobserver CT measurements and 65% of interobserver CT measurements differed by <2 mm.

CONCLUSIONS: CT-based measurements of aneurysm size tend to be larger than the US-AP measurement. CT-PCLF diameters are consistently larger than CT-PME as well as CT-AP measurements. These differences should be considered when applying evidence from previous trials to clinical decisions.

Original languageEnglish
JournalJournal of Vascular Surgery
Volume50
Issue number2
Pages (from-to)263-8
Number of pages6
ISSN0741-5214
DOIs
Publication statusPublished - Aug 2009
Externally publishedYes

Keywords

  • Aged
  • Aortic Aneurysm, Abdominal/diagnostic imaging
  • Female
  • Humans
  • Imaging, Three-Dimensional
  • Linear Models
  • Male
  • Radiographic Image Interpretation, Computer-Assisted
  • Statistics, Nonparametric
  • Tomography, X-Ray Computed/methods
  • Ultrasonography