T1 - When operable patients become inoperable
T2 - conversion of a surgical aortic valve replacement into transcatheter aortic valve implantation
AU - Olsen, Lene Kjaer
AU - Arendrup, Henrik
AU - Engstrøm, Thomas
AU - Søndergaard, Lars
PY - 2009/11
Y1 - 2009/11
N2 - Transcatheter aortic valve implantation (TAVI) is a relatively new treatment option for inoperable patients with severe aortic stenosis (AS). This case describes how a planned conventional surgical aortic valve replacement (AVR) on a 73-year-old woman was successfully converted to a TAVI procedure. On extracorporal circulation it was reconized that the aortic annulus, the coronary ostiae and the proximal part of the ascending aorta were severely calcified making valve implantation impossible. Surgical closure without valve substitution was estimated to be associated with a high risk of mortality due to the imparied left ventricular function. Consequently, TAVI was performed with a CoreValve ReValving System prosthesis. The delivery of the valve prosthesis was made through the ascending part of aorta, proximal of the cannulation of aorta. Positioning of the valve prosthesis was made under visual guidance, and the prosthesis was sutured to the ascending aorta. With some manipulation of the prosthesis it was possible to suture the aorta circumferentially around the fully expanded upper part of the prosthesis. Post-procedurally the patient recovered successfully, with improved function capacity, aortic valve area and left ventricle function.
AB - Transcatheter aortic valve implantation (TAVI) is a relatively new treatment option for inoperable patients with severe aortic stenosis (AS). This case describes how a planned conventional surgical aortic valve replacement (AVR) on a 73-year-old woman was successfully converted to a TAVI procedure. On extracorporal circulation it was reconized that the aortic annulus, the coronary ostiae and the proximal part of the ascending aorta were severely calcified making valve implantation impossible. Surgical closure without valve substitution was estimated to be associated with a high risk of mortality due to the imparied left ventricular function. Consequently, TAVI was performed with a CoreValve ReValving System prosthesis. The delivery of the valve prosthesis was made through the ascending part of aorta, proximal of the cannulation of aorta. Positioning of the valve prosthesis was made under visual guidance, and the prosthesis was sutured to the ascending aorta. With some manipulation of the prosthesis it was possible to suture the aorta circumferentially around the fully expanded upper part of the prosthesis. Post-procedurally the patient recovered successfully, with improved function capacity, aortic valve area and left ventricle function.
KW - Aged
KW - Aortic Valve/diagnostic imaging
KW - Aortic Valve Stenosis/diagnostic imaging
KW - Calcinosis/diagnostic imaging
KW - Cardiac Catheterization/instrumentation
KW - Extracorporeal Circulation
KW - Female
KW - Heart Arrest, Induced
KW - Heart Valve Prosthesis
KW - Heart Valve Prosthesis Implantation/instrumentation
KW - Humans
KW - Prosthesis Design
KW - Severity of Illness Index
KW - Suture Techniques
KW - Tomography, X-Ray Computed
KW - Treatment Outcome
U2 - 10.1510/icvts.2009.211144
DO - 10.1510/icvts.2009.211144
M3 - Journal article
C2 - 19720660
SN - 1569-9293
VL - 9
SP - 837
EP - 839
JO - Interactive Cardiovascular and Thoracic Surgery
JF - Interactive Cardiovascular and Thoracic Surgery
IS - 5
ER -