TY - JOUR
T1 - Very Early Loading of a Bone-Anchored Hearing System Is Safe
AU - Møller, Marie Louise
AU - Foghsgaard, Søren
AU - Hougaard, Dan Dupont
N1 - Copyright © 2024, Otology & Neurotology, Inc.
PY - 2024
Y1 - 2024
N2 - OBJECTIVE: To investigate the feasibility of early sound processor loading with two separate bone-anchored hearing aid implants 1 to 2 weeks after implantation. Secondary endpoints included registration of any postoperative complications, skin reactions, and any adverse events.STUDY DESIGN: Prospective, nonrandomized, noncontrolled, multicenter clinical trial.SETTING: Tertiary referral center.PATIENTS: Sixty adult patients, eligible for bone-anchored hearing aid surgery, were enrolled.INTERVENTION: Implantation of the bone-anchored hearing aid implants were done using minimally invasive surgery with tissue preservation.MAIN OUTCOME MEASURE: Implant stability quotient (ISQ) values were recorded using resonance frequency analysis. Follow-ups were scheduled 5 to 12 days, 4 to 8 weeks, 6 months, and 12 months postoperatively. Differences between the lowest ISQ values (ISQ low) over time were analyzed by paired-samples t test with a significance level of 0.05. Skin and soft tissue reactions were assessed according to Holger's classification.RESULTS: Loading time of the Ponto Wide implant was 11.6 ± 1.4 days (SD) in average (median, 12 d; range, 7-19 d). Loading of the Ponto BHX implant was done 9.7 ± 3.5 days (SD) postoperatively (median, 9 d; range, 5-19 d). The implant stability increased significantly during the 12-month study period with both implants. One Ponto BHX implant was lost. Skin and soft tissue reactions were few in both groups. Across all planned follow-up visits, no skin and soft tissue reactions were observed in 86% (Ponto BHX implant) and 85% (Ponto Wide implant).CONCLUSION: Very early loading of sound processors after percutaneous bone-anchored hearing system surgery is safe already 1 to 2 weeks postoperatively.
AB - OBJECTIVE: To investigate the feasibility of early sound processor loading with two separate bone-anchored hearing aid implants 1 to 2 weeks after implantation. Secondary endpoints included registration of any postoperative complications, skin reactions, and any adverse events.STUDY DESIGN: Prospective, nonrandomized, noncontrolled, multicenter clinical trial.SETTING: Tertiary referral center.PATIENTS: Sixty adult patients, eligible for bone-anchored hearing aid surgery, were enrolled.INTERVENTION: Implantation of the bone-anchored hearing aid implants were done using minimally invasive surgery with tissue preservation.MAIN OUTCOME MEASURE: Implant stability quotient (ISQ) values were recorded using resonance frequency analysis. Follow-ups were scheduled 5 to 12 days, 4 to 8 weeks, 6 months, and 12 months postoperatively. Differences between the lowest ISQ values (ISQ low) over time were analyzed by paired-samples t test with a significance level of 0.05. Skin and soft tissue reactions were assessed according to Holger's classification.RESULTS: Loading time of the Ponto Wide implant was 11.6 ± 1.4 days (SD) in average (median, 12 d; range, 7-19 d). Loading of the Ponto BHX implant was done 9.7 ± 3.5 days (SD) postoperatively (median, 9 d; range, 5-19 d). The implant stability increased significantly during the 12-month study period with both implants. One Ponto BHX implant was lost. Skin and soft tissue reactions were few in both groups. Across all planned follow-up visits, no skin and soft tissue reactions were observed in 86% (Ponto BHX implant) and 85% (Ponto Wide implant).CONCLUSION: Very early loading of sound processors after percutaneous bone-anchored hearing system surgery is safe already 1 to 2 weeks postoperatively.
KW - Adult
KW - Aged
KW - Aged, 80 and over
KW - Bone-Anchored Prosthesis
KW - Female
KW - Hearing Aids
KW - Humans
KW - Male
KW - Middle Aged
KW - Prospective Studies
KW - Prosthesis Implantation/methods
KW - Suture Anchors
KW - Treatment Outcome
KW - Young Adult
UR - http://www.scopus.com/inward/record.url?scp=85193112764&partnerID=8YFLogxK
U2 - 10.1097/MAO.0000000000004190
DO - 10.1097/MAO.0000000000004190
M3 - Journal article
C2 - 38693091
SN - 1531-7129
VL - 45
SP - 556
EP - 563
JO - Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology
JF - Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology
IS - 5
ER -