Validity and reliability of 3D US for the detection of erosions in patients with rheumatoid arthritis using MRI as the gold standard

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Elastography in Breast Imaging

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Simulation-Based Training of Ultrasound-Guided Procedures in Radiology - A Systematic Review

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. The EFSUMB Guidelines and Recommendations for the Clinical Practice of Elastography in Non-Hepatic Applications: Update 2018

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. The Use of Handheld Ultrasound Devices - An EFSUMB Position Paper

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  5. When Are Trainees Ready to Perform Transvaginal Ultrasound? An Observational Study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Early development of tendinopathy in humans: Sequence of pathological changes in structure and tissue turnover signaling

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Reliability of the Copenhagen Achilles length measure (CALM) on patients with an Achilles tendon rupture

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Grading lumbar disc degeneration: a comparison between low- and high-field MRI

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

PURPOSE: To test the reliability and validity of a 3D US erosion score in RA using MRI as the gold standard.

MATERIALS AND METHODS: RA patients were examined with 3D US and 3 T MRI over the 2nd and 3rd metacarpophalangeal joints. 3D blocks were evaluated by two investigators. The erosions were estimated according to a semi-quantitative score (SQS) (0 - 3) and a quantitative score (QS) (mm²). MRI was evaluated according to the RAMRIS score. For the estimation of reliability, intra-class correlation coefficients (ICC) were used. Validity was tested using Spearman's rho (rs). The sensitivity and specificity were also calculated.

RESULTS: 28 patients with RA were included. The ICC for the inter-observer reliability in the QS was 0.41 and 0.13 for the metacarpal bone and phalangeal bone, respectively, and 0.86 and 0.16, respectively, in the SQS.  The ICC for the intra-observer reliability in the QS was 0.75 and 0.48 for the metacarpal bone and phalangeal bone, respectively, and 0.83 and 0.60, respectively, in the SQS.  The correlation with MRI for the metacarpal bone was significant, with values of 0.73 (p = 0.0001) (SQ) and 0.74 (p = 0.0001) (SQS). For the phalangeal bone, bad correlation was found: 0.28 (p = 0.145) (SQ) and 0.26 (p = 0.57) (SQS). The sensitivity and specificity for the metacarpal bone were 86 % and 85 % respectively. For the phalangeal bone they were 60 % and 97 %, respectively.

CONCLUSION: Good inter- and intra-observer reliability and correlation with MRI were seen in the assessment of erosions with 3D US in the metacarpal bone, while the results were low and insignificant for the phalangeal bone, indicating that 3D US still has room for improvement.

OriginalsprogEngelsk
TidsskriftUltraschall in der Medizin (Stuttgart, Germany : 1980)
Vol/bind35
Udgave nummer2
Sider (fra-til)137-41
Antal sider5
ISSN0172-4614
DOI
StatusUdgivet - apr. 2014

ID: 44650193