Abstract
Background Ropivacaine is a widely used local anaesthetic drug, highly bound to plasma proteins with a free plasma fraction of about 5%. Therefore, the monitoring of free drug concentration is most relevant to perform pharmacokinetic studies and to understand the drug pharmacokinetic/pharmacodynamic (PK/PD) relationship. Methods A high-sensitivity liquid chromatography-tandem mass spectrometry (LC-MS/MS) method using reverse-phase LC and electrospray ionisation mass spectrometry with multiple reaction monitoring (MRM) is described for the quantitation of both free and total ropivacaine in human plasma. Ropivacaine-d7 was used as an internal standard (IS). Results The method was validated in the range 0.5-3000 ng/mL, with five levels of QC samples and according to the European Medicine Agency and Food and Drug Administration guidelines. The performance of the method was excellent with a precision in the range 6.2%-14.7%, an accuracy between 93.6% and 113.7% and a coefficient of variation (CV) of the IS-normalised matrix factor below 15%. This suitability of the method for the quantification of free and total ropivacaine in clinical samples was demonstrated with the analysis of samples from patients undergoing knee arthroplasty and receiving a local ropivacaine infiltration. Conclusions A method was developed and validated for the quantification of free and total ropivacaine in human plasma and was shown suitable for the analysis of clinical samples.