TY - JOUR
T1 - Thirty-day clinical outcomes of the Evolut PRO self-expanding transcatheter aortic valve
T2 - the international FORWARD PRO study
AU - Manoharan, Ganesh
AU - Grube, Eberhard
AU - Van Mieghem, Nicolas M
AU - Brecker, Stephen
AU - Fiorina, Claudia
AU - Kornowski, Ran
AU - Danenberg, Haim
AU - Ruge, Hendrik
AU - Thiele, Holger
AU - Lancellotti, Patrizio
AU - Søndergaard, Lars
AU - Tamburino, Corrado
AU - Oh, Jae K
AU - Fan, Yunhua
AU - Windecker, Stephan
PY - 2020/11/20
Y1 - 2020/11/20
N2 - AIMS: The Evolut PRO is a new transcatheter heart valve with an outer pericardial wrap intended to reduce paravalvular leak and facilitate tissue ingrowth. We aimed to evaluate the clinical performance and safety of the Evolut PRO valve in standard practice.METHODS AND RESULTS: FORWARD PRO is a prospective, multinational, multicentre observational study. Transcatheter aortic valve implantation with the Evolut PRO valve (23, 26, or 29 mm) was attempted in 629 non-consecutive patients from 39 centres from February 2018 to January 2019. The primary endpoint was the rate of all-cause mortality at 30 days compared to a pre-specified performance goal. An independent clinical events committee adjudicated safety endpoints based on VARC-2 definitions. All echocardiograms were centrally assessed by an independent core laboratory (Mayo Clinic, Rochester, MN, USA). Baseline characteristics included mean age 81.7±6.1 years, 61.8% female, STS score 4.7±3.3%, and 33.6% were frail. All-cause mortality at 30 days was 3.2%, which was lower than the pre-specified performance goal of 5.5% (p=0.004). Greater than mild AR was present in 1.8% of patients at discharge.CONCLUSIONS: The FORWARD PRO study confirmed the safety and efficacy of the Evolut PRO transcatheter aortic valve system with an external pericardial wrap. ClinicalTrials.gov Identifier: NCT03417011 Visual summary. Paravalvular leak at discharge and (right) the Evolut PRO valve (Medtronic) with external pericardial wrap. Reproduced with permission from Medtronic, Inc.
AB - AIMS: The Evolut PRO is a new transcatheter heart valve with an outer pericardial wrap intended to reduce paravalvular leak and facilitate tissue ingrowth. We aimed to evaluate the clinical performance and safety of the Evolut PRO valve in standard practice.METHODS AND RESULTS: FORWARD PRO is a prospective, multinational, multicentre observational study. Transcatheter aortic valve implantation with the Evolut PRO valve (23, 26, or 29 mm) was attempted in 629 non-consecutive patients from 39 centres from February 2018 to January 2019. The primary endpoint was the rate of all-cause mortality at 30 days compared to a pre-specified performance goal. An independent clinical events committee adjudicated safety endpoints based on VARC-2 definitions. All echocardiograms were centrally assessed by an independent core laboratory (Mayo Clinic, Rochester, MN, USA). Baseline characteristics included mean age 81.7±6.1 years, 61.8% female, STS score 4.7±3.3%, and 33.6% were frail. All-cause mortality at 30 days was 3.2%, which was lower than the pre-specified performance goal of 5.5% (p=0.004). Greater than mild AR was present in 1.8% of patients at discharge.CONCLUSIONS: The FORWARD PRO study confirmed the safety and efficacy of the Evolut PRO transcatheter aortic valve system with an external pericardial wrap. ClinicalTrials.gov Identifier: NCT03417011 Visual summary. Paravalvular leak at discharge and (right) the Evolut PRO valve (Medtronic) with external pericardial wrap. Reproduced with permission from Medtronic, Inc.
KW - Aged
KW - Aged, 80 and over
KW - Aortic Valve/diagnostic imaging
KW - Aortic Valve Stenosis/diagnostic imaging
KW - Female
KW - Heart Valve Prosthesis
KW - Humans
KW - Male
KW - Postoperative Complications
KW - Prospective Studies
KW - Prosthesis Design
KW - Transcatheter Aortic Valve Replacement/adverse effects
KW - Treatment Outcome
U2 - 10.4244/EIJ-D-20-00279
DO - 10.4244/EIJ-D-20-00279
M3 - Journal article
C2 - 32748789
SN - 1774-024X
VL - 16
SP - 850
EP - 857
JO - EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
JF - EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
IS - 10
ER -