Thin- versus thick-strut polymer-free biolimus-eluting stents: the BioFreedom QCA randomised trial

Manel Sabaté, Lisette Okkels Jensen, Hans-Henrik Tilsted, Raúl Moreno, Bruno García Del Blanco, Carlos Macaya, Armando Pérez de Prado, Àngel Cequier, Pedro Pérez-Fuentes, Diana Schütte, Ricardo A Costa, Hans-Peter Stoll, Jens Flensted Lassen

3 Citationer (Scopus)

Abstrakt

BACKGROUND: The BioFreedom drug-coated stent with a stainless steel platform (BF-SS) has been demonstrated to be efficacious in patients at high bleeding risk and receiv-ing only one-month dual antiplatelet therapy.

AIMS: The aim of this study was to evaluate the efficacy of the new BioFreedom Ultra drug-coated stent with a thin-strut cobalt-chromium platform (BF-CoCr) compared to the BF-SS in an all-comers population undergoing percutaneous coronary intervention (PCI).

METHODS: This was a prospective, multicentre, non-inferiority trial. The primary endpoint was in-stent late lumen loss (LLL) as determined by quantitative coronary angiography at nine-month follow-up. Clinical evaluation was performed at one year.

RESULTS: A total of 200 patients were randomised (1:1) to either the BF-CoCr or the BF-SS stent at eight centres in Spain and Denmark. Baseline clinical and lesion characteristics were similar between the groups. Mean age was 66 years and 23% were female. The mean number of stents implanted per patient was 1.5. At nine-month follow-up, mean in-stent LLL was 0.34±0.49 mm in the BF-CoCr group versus 0.29±0.37 mm in the BF-SS group, p=0.005 for non-inferiority. At one year, target lesion failure was similar between the groups (7.3% in BF-CoCr vs 9.3% in the BF-SS group; p=0.60).

CONCLUSIONS: The BF-CoCr was non-inferior to the BF-SS in terms of in-stent LLL at nine months. Larger studies powered for clinical endpoints are warranted to compare the efficacy of this new platform with currently available DES.

OriginalsprogEngelsk
TidsskriftEuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
Vol/bind17
Udgave nummer3
Sider (fra-til)233-239
Antal sider7
ISSN1774-024X
DOI
StatusUdgivet - 25 jun. 2021

Fingeraftryk

Dyk ned i forskningsemnerne om 'Thin- versus thick-strut polymer-free biolimus-eluting stents: the BioFreedom QCA randomised trial'. Sammen danner de et unikt fingeraftryk.

Citationsformater