Abstract
INTRODUCTION: The MADIT-RIT trial demonstrated reduction of inappropriate and appropriate ICD therapies and mortality by high-rate cut-off and 60-second-delayed VT therapy ICD programming in patients with a primary prophylactic ICD indication. The aim of this analysis was to study effects of MADIT-RIT ICD programming in patients with ischemic and nonischemic cardiomyopathy.
METHODS AND RESULTS: First and total occurrences of both inappropriate and appropriate ICD therapies were analyzed by multivariate Cox models in 791 (53%) patients with ischemic and 707 (47%) patients with nonischemic cardiomyopathy. Patients with ischemic and nonischemic cardiomyopathy had similar incidence of first inappropriate (9% and 11%, P = 0.21) and first appropriate ICD therapy (11.6% and 14.1%, P = 0.15). Patients with ischemic cardiomyopathy had higher mortality rate (6.1% vs. 3.3%, P = 0.01). MADIT-RIT high-rate cut-off (arm B) and delayed VT therapy ICD programming (arm C) compared with conventional (arm A) ICD programming were associated with a significant risk reduction of first inappropriate and appropriate ICD therapy in patients with ischemic and nonischemic cardiomyopathy (HR range 0.11-0.34, P < 0.001 for all comparisons). Occurrence of total inappropriate and appropriate ICD therapies was significantly reduced by high-rate cut-off ICD programming and delayed VT therapy ICD programming in both ischemic and nonischemic cardiomyopathy patients.
CONCLUSION: High-rate cut-off and delayed VT therapy ICD programming are associated with significant reduction in first and total inappropriate and appropriate ICD therapy in patients with ischemic and nonischemic cardiomyopathy.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Journal of Cardiovascular Electrophysiology |
Vol/bind | 26 |
Udgave nummer | 4 |
Sider (fra-til) | 424-33 |
Antal sider | 10 |
ISSN | 1045-3873 |
DOI | |
Status | Udgivet - apr. 2015 |