Surveillance for hemodialysis access stenosis: usefulness of ultrasound vector volume flow

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Rating scale for the assessment of competence in ultrasound-guided peripheral vascular access - a Delphi Consensus Study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Current use of ultrasound for central vascular access in children and infants in the Nordic countries--a cross-sectional study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Blood flow measurements during hemodialysis vascular access interventions - Catheter-based thermodilution or Doppler ultrasound?

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Catheter-based flow measurements in hemodialysis fistulas - Bench testing and clinical performance

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  5. Endovascular treatment of hemodialysis arteriovenous fistulas: is immediate post-interventional blood flow a predictor of patency

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

PURPOSE: To investigate if ultrasound vector-flow imaging (VFI) is equal to the reference method ultrasound dilution technique (UDT) in estimating volume flow and changes over time in arteriovenous fistulas (AVFs) for hemodialysis.

MATERIALS AND METHODS: From January 2014 to January 2015, patients with end-stage renal disease and matured functional AVFs were consecutively solicited to participate in this prospective study. All patients were included after written informed consent and approval by the National Committee on Biomedical Research Ethics and the local Ethics Committee (journal no. H-4-2014-FSP). VFI and UDT measurements were performed monthly over a six-month period. Nineteen patients were included in the study. VFI measurements were performed before dialysis, and UDT measurements after. Statistical analyses were performed with Bland-Altman plot, Student's t-test, four-quadrant plot, and regression analysis. Repeated measurements and precision analysis were used for reproducibility determination.

RESULTS: Precision measurements for UDT and VFI were 32% and 20%, respectively (p = 0.33). Average volume flow measured with UDT and VFI were 1161 mL/min (±778 mL/min) and 1213 mL/min (±980 mL/(min), respectively (p = 0.3). The mean difference was -51 mL/min (CI: -150 mL/min to 46 mL/min) with limits of agreement from -35% to 54%, with a strong correlation (r2 = 0.87). A large change in volume flow between dialysis sessions detected by UDT was confirmed by VFI (p = 0.0001), but the concordance rate was poor (0.72).

CONCLUSIONS: VFI is an acceptable method for volume flow estimation and volume flow changes over time in AVFs.

OriginalsprogEngelsk
TidsskriftThe journal of vascular access
Vol/bind17
Udgave nummer6
Sider (fra-til)483-488
Antal sider6
ISSN1129-7298
DOI
StatusUdgivet - 2 nov. 2016

ID: 49464107