Stroke outcomes following durable left ventricular assist device implant in patients bridged with micro-axial flow pump: Insights from a large registry

Guglielmo Gallone*, Daniel Lewin, Sebastian Rojas Hernandez, Alexander Bernhardt, Michael Billion, Anna Meyer, Ivan Netuka, J-J Kooij, Marina Pieri, Mariusz K Szymanski, Christian H Moeller, Payam Akhyari, Khalil Jawad, Ihor Krasivskyi, Bastian Schmack, Gloria Färber, Marta Medina, Assad Haneya, Daniel Zimpfer, Gaik NersesianPia Lanmueller, Antonio Spitaleri, Mehmet Oezkur, Ilija Djordjevic, Diyar Saeed, Massimo Boffini, Julia Stein, F Gustafsson, Anna Mara Scandroglio, Gaetano Maria De Ferrari, Bart Meyns, Steffen Hofmann, Jan Belohlavek, Jan Gummert, Mauro Rinaldi, Evgenij V Potapov, Antonio Loforte

*Corresponding author af dette arbejde
3 Citationer (Scopus)

Abstract

BACKGROUND: Stroke after durable left ventricular assist device (d-LVAD) implantation portends high mortality. The incidence of ischemic and hemorrhagic stroke and the impact on stroke outcomes of temporary mechanical circulatory support (tMCS) management among patients requiring bridge to d-LVAD with micro-axial flow-pump (mAFP, Abiomed) is unsettled.

METHODS: Consecutive patients, who underwent d-LVAD implantation after being bridged with mAFP at 19 institutions, were retrospectively included. The incidence of early ischemic and hemorrhagic stroke after d-LVAD implantation (<60 days) and association of pre-d-LVAD characteristics and peri-procedural management with a specific focus on tMCS strategies were studied.

RESULTS: Among 341 patients, who underwent d-LVAD implantation after mAFP implantation (male gender 83.6%, age 58 [48-65] years, mAFP 5.0/5.5 72.4%), the early ischemic stroke incidence was 10.8% and early hemorrhagic stroke 2.9%. The tMCS characteristics (type of mAFP device and access, support duration, upgrade from intra-aortic balloon pump, ECMELLA, ECMELLA at d-LVAD implantation, hemolysis, and bleeding) were not associated with ischemic stroke after d-LVAD implant. Conversely, the device model (mAFP 2.5/CP vs. mAFP 5.0/5.5: HR 5.6, 95%CI 1.4-22.7, p = 0.015), hemolysis on mAFP support (HR 10.5, 95% CI 1.3-85.3, p = 0.028) and ECMELLA at d-LVAD implantation (HR 5.0, 95% CI 1.4-18.7, p = 0.016) were associated with increased risk of hemorrhagic stroke after d-LVAD implantation. Both early ischemic (HR 2.7, 95% CI 1.9-4.5, p < 0.001) and hemorrhagic (HR 3.43, 95% CI 1.49-7.88, p = 0.004) stroke were associated with increased 1-year mortality.

CONCLUSIONS: Among patients undergoing d-LVAD implantation following mAFP support, tMCS characteristics do not impact ischemic stroke occurrence, while several factors are associated with hemorrhagic stroke suggesting a proactive treatment target to reduce this complication.

OriginalsprogEngelsk
TidsskriftArtificial Organs
Vol/bind48
Udgave nummer10
Sider (fra-til)1168-1179
Antal sider12
ISSN0160-564X
DOI
StatusUdgivet - okt. 2024

Fingeraftryk

Dyk ned i forskningsemnerne om 'Stroke outcomes following durable left ventricular assist device implant in patients bridged with micro-axial flow pump: Insights from a large registry'. Sammen danner de et unikt fingeraftryk.