Abstract
PURPOSE: To present a modification of the Zenith TX2 thoracic stent-graft that allows staged proximal deployment to improve apposition to the aortic wall.
TECHNIQUE: Three standard Zenith TX2 thoracic stent-grafts and 3 modified versions were deployed in a glass model of the aortic arch. Deployment sequences were analyzed. In a patient with a 6-cm thoracic aortic aneurysm after a type B dissection, the modified Zenith TX2 thoracic stent-graft was deployed successfully and without complications; the proximal part of the stent-graft protruded less into the arch, significantly improving wall apposition.
CONCLUSION: Staged proximal deployment with a modified Zenith TX2 thoracic stent-graft can improve orientation and wall apposition of the first Z stent. A better proximal apposition may prevent early and late stent-graft complications.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists |
Vol/bind | 16 |
Udgave nummer | 5 |
Sider (fra-til) | 598-602 |
Antal sider | 5 |
ISSN | 1526-6028 |
DOI | |
Status | Udgivet - okt. 2009 |
Udgivet eksternt | Ja |