Spring til hovednavigation Spring til søgning Spring til hovedindhold

Secondary knee instability caused by fracture of the stabilizing insert in a dual-articular total knee

Morten P Boesen, Tim Toftgaard Jensen, Henrik Husted

8 Citationer (Scopus)

Abstract

A case of a fractured polyethylene stabilizing insert causing secondary knee instability in a Dual-articular total knee arthroplasty (TKA) is presented. A 65-year-old woman who underwent surgery with a Dual-articular TKA 4 years earlier had a well-functioning prosthesis until a fall, after which she began complaining of an intermittent, audible clicking; pain; and instability. At surgery, a fractured polyethylene peg was discovered, and the tibial insert was replaced. At follow-up, she remains asymptomatic. This is the first description of a fractured stabilizing insert in a Dual-articular knee.
OriginalsprogEngelsk
TidsskriftJournal of Arthroplasty
Vol/bind19
Udgave nummer7
Sider (fra-til)941-943
Antal sider3
ISSN0883-5403
StatusUdgivet - 2004