TY - JOUR
T1 - Risk factors for cardiac implantable electronic device infections
T2 - a nationwide Danish study
AU - Olsen, Thomas
AU - Jørgensen, Ole Dan
AU - Nielsen, Jens Cosedis
AU - Thøgersen, Anna Margrethe
AU - Philbert, Berit Thornvig
AU - Frausing, Maria Hee Jung Park
AU - Sandgaard, Niels Christian Foldager
AU - Johansen, Jens Brock
N1 - © The Author(s) 2022. Published by Oxford University Press on behalf of the European Society of Cardiology.
PY - 2022/12/14
Y1 - 2022/12/14
N2 - AIMS: Cardiac implantable electronic device (CIED) infection is a severe complication to modern management of cardiac arrhythmias. The CIED type and the type of surgery are recognized as risk factors for CIED infections, but knowledge of patient-related risk factors is scarce. This study aimed to identify lifelong patient-related risk factors for CIED infections.METHODS AND RESULTS: Consecutive Danish patients undergoing a CIED implantation or reoperation between January 1996 and April 2018 were included. The cohort consisted of 84 429 patients undergoing 108 494 CIED surgeries with a combined follow-up of 458 257 CIED-years. A total of 1556 CIED explantations were classified as either pocket (n = 1022) or systemic CIED infection (n = 534). Data were cross-linked with records from the Danish National Patient Registry and the Danish National Prescription Registry. Using multiple-record and multiple-event per subject proportional hazard analysis, specific patient-related risk factors were identified but with several variations amongst the subtypes of CIED infection. CIED reoperations were associated with the highest risk of pocket CIED infection but also CIED type, young age, and prior valvular surgery [hazard ratio (HR): 1.62, 95% confidence interval (CI): 1.29-2.04]. Severe renal insufficiency/dialysis (HR: 2.40, 95% CI: 1.65-3.49), dermatitis (HR: 2.80, 95% CI: 1.92-4.05), and prior valvular surgery (HR: 2.09, 95% CI: 1.59-2.75) were associated with the highest risk of systemic CIED infections. Congestive heart failure, ischaemic heart disease, malignancy, chronic obstructive pulmonary disease, and temporary pacing were not significant at multivariate analysis.CONCLUSION: Specific comorbidities and surgical procedures were associated with a higher risk of CIED infections but with variations amongst pocket and systemic CIED infection. Pocket CIED infections were associated with CIED reoperations, young age and more complex type of CIED, whereas systemic CIED infections were associated with risk factors predisposing to bacteraemia.
AB - AIMS: Cardiac implantable electronic device (CIED) infection is a severe complication to modern management of cardiac arrhythmias. The CIED type and the type of surgery are recognized as risk factors for CIED infections, but knowledge of patient-related risk factors is scarce. This study aimed to identify lifelong patient-related risk factors for CIED infections.METHODS AND RESULTS: Consecutive Danish patients undergoing a CIED implantation or reoperation between January 1996 and April 2018 were included. The cohort consisted of 84 429 patients undergoing 108 494 CIED surgeries with a combined follow-up of 458 257 CIED-years. A total of 1556 CIED explantations were classified as either pocket (n = 1022) or systemic CIED infection (n = 534). Data were cross-linked with records from the Danish National Patient Registry and the Danish National Prescription Registry. Using multiple-record and multiple-event per subject proportional hazard analysis, specific patient-related risk factors were identified but with several variations amongst the subtypes of CIED infection. CIED reoperations were associated with the highest risk of pocket CIED infection but also CIED type, young age, and prior valvular surgery [hazard ratio (HR): 1.62, 95% confidence interval (CI): 1.29-2.04]. Severe renal insufficiency/dialysis (HR: 2.40, 95% CI: 1.65-3.49), dermatitis (HR: 2.80, 95% CI: 1.92-4.05), and prior valvular surgery (HR: 2.09, 95% CI: 1.59-2.75) were associated with the highest risk of systemic CIED infections. Congestive heart failure, ischaemic heart disease, malignancy, chronic obstructive pulmonary disease, and temporary pacing were not significant at multivariate analysis.CONCLUSION: Specific comorbidities and surgical procedures were associated with a higher risk of CIED infections but with variations amongst pocket and systemic CIED infection. Pocket CIED infections were associated with CIED reoperations, young age and more complex type of CIED, whereas systemic CIED infections were associated with risk factors predisposing to bacteraemia.
KW - Defibrillators, Implantable/adverse effects
KW - Denmark/epidemiology
KW - Electronics
KW - Humans
KW - Incidence
KW - Pacemaker, Artificial/adverse effects
KW - Prosthesis-Related Infections/epidemiology
KW - Retrospective Studies
KW - Risk Factors
UR - http://www.scopus.com/inward/record.url?scp=85144584228&partnerID=8YFLogxK
U2 - 10.1093/eurheartj/ehac576
DO - 10.1093/eurheartj/ehac576
M3 - Journal article
C2 - 36263789
SN - 0195-668X
VL - 43
SP - 4946
EP - 4956
JO - European Heart Journal
JF - European Heart Journal
IS - 47
ER -